    Assessment the effect of CT localizer parameter on radiation dose and image quality of GE CT spiral images, A phantom study

    (ندگان)پدیدآور
    Salimi, YazdanDeevband, MohammadRezaGhafarian, Pardis
    Thumbnail
    نوع مدرک
    Text
    Conference Proceedings
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: Using computed tomography (CT) is increasing significantly in Iran and worldwide. Reduction of CT's radiation dose is a critical concern. The aim of this study was evaluating the effect of localizer kVp on spiral CT images when the tube current modulation (TCM) is activated. Materials and Methods: A special wedding cake phantom with circular intersection and diameters of 15 to 40 cm was scanned on a GE VCT 64 slice scanner. The whole length of phantom (84cm) was scanned with kVps of 80,100,120 and 140. The image noise was measured in SD of Ct numbers and the radiation dose was measured in term of CTDIvol. The tube current variation through the Z-Axis was extracted from DICOM images. Results: Function of TCM was affected by changing the scout view parameters. By increasing the localizer kVp the average used tube current was decreased and image noise was increased (8-32%, pConclusion: The attenuation of localizer X-ray is the criteria of tube current changing in spiral CT. The attenuation is variable in different kVps of X-ray. The ability of TCM system was the best in offering the constant image noise when the localizer and spiral scans were in the same kVp. Inappropriate localizer kVp may increase the patients' radiation dose or degrade the image quality by miscalculations in TCM system
    کلید واژگان
    Computed Tomography Tube Current Modulation Patient Dose
    kVp

    شماره نشریه
    12
    تاریخ نشر
    2018-12-01
    1397-09-10
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Department of biomedical engineering and Medical Physics, Faculty of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran
    Department of biomedical engineering and Medical Physics, Faculty of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran
    Chronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases (NRITLD), Shahid Beheshti University of Medical Sciences, Tehran, Iran. PET/CT and Cyclotron Center of Masih Daneshvari Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2018.12977
    http://ijmp.mums.ac.ir/article_12977.html
    https://iranjournals.nlai.ir/handle/123456789/324610

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب