      Standardization and Validation of Non-invasive Monitoring of End Tidal Carbon Dioxide in Neonates via Nasal Cannula: An Observational Study

      (ندگان)پدیدآور
      Nayak, Varun RAnn Skariah, TishaLewis, Leslie
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      770.5کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Article
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Background: Carbon dioxide (CO2) is a by-product of cellular metabolism, which could be considered as a reflection of metabolism, circulation, and ventilation. Arterial blood gas analysis (ABG) is the gold standard of monitoring for CO2. However, is an expensive method leading to blood loss and iatrogenic anemia. In addition, each sample is only a snapshot view of the sampling moment. End-tidal CO2 (ETCO2) measurement gives a non-invasive and continuous monitoring of exhaled CO2. Therefore, this study aimed to validate the non-invasive CO2 measurement performed by nasal cannula and evaluate the correlation with partial pressure of arterial CO2 (PaCO2 ) in neonates. Methods: This single-center observational study was conducted in the Neonatal Intensive Care Unit (NICU), Kasturba Hospital, Manipal, India. PaCO2 was reported on routine ABG within a 15-minute interval of ABG sampling. Moreover, partial pressure of end-tidal carbon dioxide (PetCO2) was noted at continuous 30-sec intervals (i.e., 30, 60, 90, ..., 180) up to 3 minutes. The values of PaCO2 and PetCO2 were found to be correlated with Pearson correlation and were shown by scattered plot. Regression analysis was used to get the prediction equation and the variance. Results: A total of 70 samples were taken in the initial phase to study the correlation between PaCO2 and measured PetCO2. Pearson correlation showed a moderate positive correlation (r= 0.589) between PetCO2 and measured PaCO2. Regression analysis demonstrated a variance of 33.8% between the measured PetCO2 and PaCO2, which was statistically significant (P2. In the final phase, 20 samples were recruited to standardize and validate the prediction equation. The PaCO2 was calculated using the predicted equation and a new prediction equation was obtained. Conclusion: According to the findings of this study, there is a good correlation (r=0.681) between the non-invasively measured PetCO2 and PaCO2.
      کلید واژگان
      correlation
      ETCO2
      Nasal cannula
      Neonates
      Non-invasive
      PaCO2

      شماره نشریه
      2
      تاریخ نشر
      2019-06-01
      1398-03-11
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Department of Respiratory Therapy, School of Allied Health Sciences, Manipal Academy of Higher Education, Manipal, Karnataka, India
      Department of Respiratory Therapy, School of Allied Health Sciences, Manipal Academy of Higher Education, Manipal, Karnataka, India
      Head of Neonatal Intensive Care Unit, Kasturba Hospital Manipal, Karnataka, India

      شاپا
      2251-7510
      2322-2158
      URI
      https://dx.doi.org/10.22038/ijn.2018.32704.1460
      http://ijn.mums.ac.ir/article_12001.html
      https://iranjournals.nlai.ir/handle/123456789/323367

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب