Comparisons between the KKU-Model and Conventional Rectal Tubes as Markers for Checking Rectal Doses during Intracavitary Brachytherapy of Cervical Cancer
(ندگان)پدیدآور
پدیدآور نامشخصنوع مدرک
Textزبان مدرک
Englishچکیده
Background: To compare the KKU-model rectal tube (KKU-tube) and the conventional rectal tube (CRT) for checking rectal doses during high-dose-rate intracavitary brachytherapy (HDR-ICBT) of cervical cancer. Materials and Methods: Between February 2010 and January 2011, thirty -two patients with cervical cancer were enrolled and treated with external beam radiotherapy (EBRT) and intracavitary brachytherapy (ICBT). The KKU-tube and CRT were applied intrarectally in the same patients at alternate sessions as references forcalculation of rectal doses during ICBT. The gold standard references of rectum anatomical markers which are most proximal to radiation sources were anterior rectal walls (ARW) adjacent to the uterine cervix demonstrated by barium sulfate suspension enema. The calculated rectal doses derived from actual anterior rectal walls, CRT and the anterior surfaces of the KKU-tubes were compared by using the paired t-test. The pain caused by insertion of each type of rectal tube was assessed by the visual analogue scale (VAS). Results: The mean dose of CRT was lower than the mean dose of ARW (Dmean0-Dmean1) by 80.55±47.33 cGy (p-value Conclusions: The KKU-model rectal tube was found to be an easily producible, applicable and reliable instrument as a reference for evaluating the rectal dose during ICBT of cervical cancer without negative effects on the patients.
کلید واژگان
KKU-model rectal tuberectal dose
intracavitary brachytherapy
cervical cancer
شماره نشریه
15تاریخ نشر
2014-12-011393-09-10
ناشر
West Asia Organization for Cancer Prevention (WAOCP)شاپا
1513-73682476-762X