      Planning and Dosimetric Study of Volumetric Modulated Arc Based Hypofractionated Stereotactic Radiotherapy for Acoustic Schwannoma - 6MV Flattening Filter Free Photon Beam

      (ندگان)پدیدآور
      پدیدآور نامشخص
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      1.577 مگابایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Background: The purpose of this study was to assess the dosimetric and clinical feasibility of volumetricmodulated arc based hypofractionated stereotactic radiotherapy (RapidArc) treatment for large acousticschwannoma (AS >10cc). Materials and 10cc). Materials and Methods: Ten AS patients were immobilized using BrainLab mask.They were subject to multimodality imaging (magnetic resonance and computed tomography) to contour targetand organs at risk (brainstem and cochlea). Volumetric modulated arc therapy (VMAT) based stereotactic planswere optimized in Eclipse (V11) treatment planning system (TPS) using progressive resolution optimizer-III andfinal dose calculations were performed using analytical anisotropic algorithm with 1.5 mm grid resolution. All ASpresented in this study were treated with VMAT based HSRT to a total dose of 25Gy in 5 fractions (5fractions/week). VMAT plan contains 2-4 non-coplanar arcs. Treatment planning was performed to achieve at least 99%of PTV volume (D99) receives 100% of prescription dose (25Gy), while dose to OAR’s were kept below thetolerance limits. Dose–volume histograms (DVH) were analyzed to assess plan quality. Treatments were deliveredusing upgraded 6 MV un-flattened photon beam (FFF) from Clinac-iX machine. Extensive pretreatment qualityassurance measurements were carried out to report on quality of delivery. Point dosimetry was performed usingthree different detectors, which includes CC13 ion-chamber, Exradin A14 ion-chamber and Exradin W1 plasticscintillator detector (PSD) which have measuring volume of 0.13 cm3, 0.009 cm3 and 0.002 cm3 respectively.Results: Average PTV volume of AS was 11.3cc (±4.8), and located in eloquent areas. VMAT plans providedcomplete PTV coverage with average conformity index of 1.06 (±0.05). OAR’s dose were kept below tolerancelimit recommend by American Association of Physicist in Medicine task group-101(brainstem V0.5cc
      کلید واژگان
      Acoustic schwannoma
      VMAT
      HSRT
      point dosimetry
      COMPASS

      شماره نشریه
      12
      تاریخ نشر
      2015-12-01
      1394-09-10
      ناشر
      West Asia Organization for Cancer Prevention (WAOCP)

      شاپا
      1513-7368
      2476-762X
      URI
      http://journal.waocp.org/article_31205.html
      https://iranjournals.nlai.ir/handle/123456789/37584

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب