Calculation of Equivalent Square Fields in Radiation Therapy by an Improved Vadash Correction Factor for Collimator Exchange Effects
(ندگان)پدیدآور
Tahmasebi Birgani, Mohammad JavadChegni, NahidTahmasbi, MarziyehFayazi, RaziyeFatahi Asl, JafarBehrouz, Mohammad Aliنوع مدرک
TextResearch Articles
زبان مدرک
Englishچکیده
Background and objectives: To determine the head scatter factor, a formalism presented by Vadash and Bjärngardhas been employed to assess collimator exchange effects. The aim of this study was to determine the best Vadashcorrection factor (A) by introducing a new method based on the output factor measured in air for different square andrectangular fields. Materials and Methods: A new simple mathematical method based on selection of the best dosimetricdata was proposed to obtain the A value for Vadash to predict the equivalent square field size. Measurements wereperformed with a Farmer c