Determination of Diazinon in Environmental Samples Using Modified Multi-Walled Carbon Nanotubes as Pipette-Tip Solid Phase Extraction Sorbent
(ندگان)پدیدآور
Rezaei Kahkha, Mohammad RezaKaykhaii, MassoudShafee-Afarani, MahdiRezaei Kahkha, Batoolنوع مدرک
TextOriginal research article
زبان مدرک
Englishچکیده
In this work, a microextraction technique based on pipette tip solid-phase extraction was used for preconcentration and determination of diazinon. Carbon nanotube functionalized by zinc sulfide and ethylene glycol was used as sorbent. Determination of diazinon was performed using high performance liquid chromatography and UV detection. Important parameters that influence the extraction efficiency (i.e. pH, amount of adsorbent, extraction time, salt addition, volumes of sample and eluting solvent and number of aspirating/dispensing cycles for both solvent and sample) were investigated and optimized. Results were showed that method was validated over the range of 0.50 - 100.0 µg L-1. Repeatability was satisfactory, bellow 3.78% for 5 replicate measurements of 20 µg L-1 of diazinon. The limit of detection of this method is 0.03 µg L-1 with an enrichment factor of 100 and short extraction time of 8.5 min, which confirmed suggested method is a reliable and accurate for extraction and preconcentration of diazinon.
کلید واژگان
Carbon NanotubesDiazinon
Pipette Tip
Preconcentration
solid phase extraction
شماره نشریه
2تاریخ نشر
2017-09-011396-06-10
ناشر
Payame Noor University, Iranدانشگاه پیام نور
سازمان پدید آورنده
Department of Environmental Health Engineering, Faculty of Health, Zabol University of Medical Sciences, Zabol, IranDepartment of Chemistry, Faculty of Sciences, University of Sistan and Baluchestan, Zahedan, Iran
Department of Material Engineering, University of Sistan and Baluchestan, Zahedan, Iran
Department of Occupational Health Engineering, Kerman University of Medical Sciences, Kerman, Iran
شاپا
2383-22072538-5054