Dosimetric Validation of Digital Megavolt Imager for Flattening Filter Free Beams in the Pre-Treatment Quality Assurance of Stereotactic Body Radiation Therapy for Liver Metastases
(ندگان)پدیدآور
Subramani, VendhanRathakrishnan, MuraliN, Arunai NambirajS, Saraswathi ChitraVenkatraman, Muraliنوع مدرک
TextResearch Articles
زبان مدرک
Englishچکیده
Aim: The aim is to evaluate the use of digital megavolt imager (DMI) aS1200 in portal dosimetry with flattening filter free (FFF) beams. Materials and Methods: Dosimetric properties of DMI is characterized at 6MV FFF beams for signal saturation, dose linearity, dependency on dose-rate and source-detector distance (SDD), signal lag (ghosting), and back scatter. Portal dosimetry is done for twenty volumetric modulated arc therapy (VMAT) based stereotactic body radiotherapy (SBRT) plans for the treatment of liver metastases and the results are compared with repeated measurements of Octavius 4D. Results: The detector signal to monitor unit (MU) ratio drops drastically below 25MU. The detector linearity with dose is within 1% and no evidence of signal saturation as such. The aS1200 response variation across various dose rates and SDD is
کلید واژگان
SBRT QALiver metastates
FFF beams
DMI detector
Portal dosimetry
Radiation oncology
شماره نشریه
6تاریخ نشر
2020-06-011399-03-12
ناشر
West Asia Organization for Cancer Prevention (WAOCP)سازمان پدید آورنده
Department of Radiation Oncology, Apollo Cancer Institute, Chennai, Tamilnadu, India.Department of Physics, School of Advanced Sciences, VIT University, Vellore, Tamilnadu, India.
Centre for Biomaterials, Cellular and Molecular Theranostics, VIT University, Vellore, Tamilnadu, India.
Department of Radiation Oncology, Apollo Cancer Institute, Chennai, Tamilnadu, India.
Department of Radiation Oncology, Apollo Cancer Institute, Chennai, Tamilnadu, India.
شاپا
1513-73682476-762X