      Dosimetric Characteristics of Transparent Bolus for External Beam Radiotherapy

      (ندگان)پدیدآور
      Govindaraj, KesavanS, SenthilKumarR.K, Jeevan Ram
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      935.8کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: In radiotherapy, the bolus is often used while treating the tumor under the uneven surfaces of the patients for correcting the anatomical irregularities and increasing skin dose. Wet cotton and wet gauze are still used in developing countries, since the use of wet cotton and wet gauze has certain disadvantages, there is a need for transparent bolus which should be similar to a universally accepted bolus in terms of properties with a lower cost (50% less expensive). Material and Methods: The present study was conducted to investigate the characteristics of transparent bolus (Senflab) material, such as transmission factor, percentage depth dose (PDD), stability over time and high dose, homogeneity and transparency using 6 and 15 MV photons beam and 12, 15, and 18 MeV electrons beam. Moreover, the new bolus material was compared with those of the commercially available Superflab and RW3 slab. Results: The percentage difference in the transmission factor of Senflab was less than ±1.9%, compared with Superflab and RW3. For PDD, the percentage difference was ±2.88% and ±1.26% for photons and electron beams, respectively. The performance of bolus remained constant both physically and dosimetrically after higher dose exposure. The percentage standard deviation was 0.0002% for a period of one month, and 0.0003% for the homogeneity. The transparency of the bolus material was good enough to display the set radiation treatment field. Conclusion: This study shows the suitability of the new bolus for routine use in radiotherapy.
      کلید واژگان
      Surface Dose
      Bolus
      Radiotherapy
      Medical Physics
      Physics of Radiotherapy
      Radiation Detection & Measurement

      شماره نشریه
      6
      تاریخ نشر
      2019-11-01
      1398-08-10
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Research and development centre,Bharathiar University,Coimbatore,India
      Dept. of Radiotherapy, Madurai Medical College & Govt. Rajaji Hospital, Madurai, India.
      Dept.of Medical Physics, Bharathidasan University, Tiruchirapalli, India

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2019.37129.1470
      http://ijmp.mums.ac.ir/article_12906.html
      https://iranjournals.nlai.ir/handle/123456789/325005

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب