      New Method of Quality Control Test for Light and Radiation Field Coincidence in Medical Linear Accelerators

      (ندگان)پدیدآور
      Heravian Shandiz, MahdiSafaeian Layen, GhorbanBahreyni-Toosi, Mohammad-HosseinFeyzi, AliZiaei, Amir Hossein
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      840.9کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: The evaluation of X-ray and light field coincidence in linear accelerators as a quality control test is often performed subjectively, involving the manual marking of films and their visual inspection following the irradiation. Therefore, the present study aimed to develop an objective method for the performance of this test leading to the increased levels of accuracy, precision, and speed for the measurement of X-ray and light field coincidence. Material and Methods: The new method involved a portable, lightweight, and inexpensive device containing optically-shielded and non-shielded photodiodes to detect the location and dimensions of the light and X-ray fields. The obtained results were analyzed using purpose-written user-friendly software. Results: On the basis of the results, this system could be a reliable method to measure the coincidence of the two fields with the accuracy of 0.5 mm and average field size standard deviations of Elekta Presice and Siemens Primus are 22.47 mm2 and 22.36 mm2, respectively. The result was well within the tolerance recommended by the American Association of Physicists in Medicine task group report number 142. Conclusion: The proposed method allows accurate and precise measurements through a largely automated process. Therefore, the measurement results benefit from the reduced level of subjectivity or human error, compared to the standard film-based technique.
      کلید واژگان
      Quality Control
      Quality Assurance
      Radiation
      Medical Physics
      Physics of Radiotherapy

      شماره نشریه
      5
      تاریخ نشر
      2019-09-01
      1398-06-10
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Biomedical Engineering Department, Imam Reza Educational, Research and Treatment Center, Mashhad University of Medical Science, Mashhad, Iran
      Department of Technology of Radiology, school of paramedical science, Mashhad University of Medical Science, Mashhad, Iran
      Medical Physics Research Center, Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran
      Department of Technology of Radiology, school of paramedical science, Mashhad university of medical science, Mashhad, Iran
      Department of Technology of Radiology, school of paramedical science, Mashhad University of Medical Science, Mashhad, Iran

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2019.21250.1202
      http://ijmp.mums.ac.ir/article_13084.html
      https://iranjournals.nlai.ir/handle/123456789/324993

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب