      In Vivo Dosimetry Using a Flat Surface Sun Nuclear Corporation Diode in 60co Beams for Some Radiotherapy Treatments in Ghana

      (ندگان)پدیدآور
      Manson, EricAtuwo‑Ampoh, VivianTagoe, SamuelFiagbedzi, EmmanuelAddison, EricSchandorf, Cyril
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      990.1کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: One of the useful standard quality assurance techniques in radiation therapy is monitoring entrance doses in in-vivo dosimetry. An overall tolerance limit of 5% of the absorbed radiation dose has been recommended by the International Commission of Radiological Units. The implementation of an in vivo dosimetry still remains as a challenge to clinical medical physicists. As a result, the practice of constant monitoring of patients undergoing radiation therapy in most of the radiotherapy departments in Africa has not been given much attention. The study aimed at the evaluation of in-vivo entrance dosimetry using diodes to verify the accuracy of the radiation delivered to patients, compared to prescribed doses. Material and Methods: In this paper, a protocol for in vivo dosimetry using a two flat surface Sun Nuclear Corporation diode in a radiotherapy department has been implemented in equinox Cobalt 60 beams. A water phantom calibrated was performed using the International Atomic Energy Agency standards (TRS 398). Calibration coefficients were determined with diodes using a Perspex phantom to derive correction factors. A total number of 137 patients' doses were measured with the diodes during the treatment of 4 different sites. Results: The average deviation between the measured and expected entrance dose performed by the phantom studies was 5% (0.34±1.8%) in almost all cases. Conclusion: The developed protocol in this study indicates that in vivo dosimetry using silicon diodes is reliable, which can be adopted as a universal quality assurance tool in the radiotherapy departments. Moreover, measurements with diodes can be acquired online which produces an instant readout and is relatively cheaper as compared to the ion chamber.
      کلید واژگان
      In Vivo Dosimetry
      Co-60
      Diode
      Radiotherapy
      Medical Physics
      Physics of Radiotherapy
      Radiation Detection & Measurement

      شماره نشریه
      5
      تاریخ نشر
      2019-09-01
      1398-06-10
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      University of Ghana Medical School
      Oncology Department Komfo Anokye Teaching Hospital Kumasi Ghana
      National Radiotherapy and Nuclear Medicine Centre, Korle-Bu Teaching Hospital, Accra-Ghana
      Oncology Department KomfoAnokye Teaching Hospital P. O. Box 1934 Kumasi – Ghana
      Physics Department Kwame Nkrumah University of Science and Technology Kumasi – Ghana
      Ghana Atomic Energy Commission and School of Nuclear and Allied Sciences, University of Ghana

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2018.29705.1324
      http://ijmp.mums.ac.ir/article_12018.html
      https://iranjournals.nlai.ir/handle/123456789/324992

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب