    Optimization of Radiation Dose in Cranial Computed Tomography among Adults: Assessment of Radiation Dose against Image Quality

    (ندگان)پدیدآور
    Okoro, NnamdiChangizi, VahidJazayeri, ElahePak, Farideh
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    1.328 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: The rapid use of computed tomography (CT) scan is of great concern, due to increase in patients' dose. Optimization of CT protocol is a vital issue in dose reduction. This study aimed to optimize radiation dose in cranial CT and assess modifications in image quality under radiation dose reduction. Material and Methods: A poly(methyl methacrylate) phantom was used for quality control test on CT scanners. Data of 214 scan parameters, dose indicators; volume CT dose index (CTDIvol) and dose-length product (DLP) of patients who underwent cranial CT scans were collected. The data were grouped into three, with respect to the slice numbers of 24, 28, and 32. Tube voltage (kVp) and slice thickness were constant; (110 kVp and 4.8 mm, respectively), at variable tube currents (mAs). A one-sample t-test was used to compare the dose indicator values of the hospital protocol with a recommended protocol. Scan parameters were optimized for radiation dose against image quality. Results: Increased mAs resulted in increased CTDIvol and DLP at constant kVp and slice thickness. Moreover, dose indicators recorded the lowest and highest values at the slice numbers of 24 and 32, respectively. An increase in slice numbers affected dose indicators. Dose indicators recorded significant reduction (PConclusion: Optimization of CT protocol considers radiation dose and image quality. Radiologists adopted protocols acquired with lower scan parameters and dose indicators lower than the recommended achievable dose limit of 58 mGy.
    کلید واژگان
    optimization
    Image Quality
    Radiation Dose
    Computed Tomography
    Computed Tomography (CT)
    Medical Physics
    Radiation Detection & Measurement

    شماره نشریه
    5
    تاریخ نشر
    2020-09-01
    1399-06-11
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Department of Technology of Radiology and Radiotherapy, School of Allied Medical Sciences, Tehran University of Medical Sciences, Tehran, Iran
    Technology of Radiology and Radiotherapy Department Allied Medical Sciences School, Tehran University of Medical Sciences, Tehran, Iran
    Technology of Radiology and Radiotherapy Department, Allied Medical Sciences School, Tehran University of Medical Sciences, Tehran, Iran
    Department of Technology of Radiology and Radiotherapy, School of Allied Medical Sciences, Tehran University of Medical Sciences, Tehran, Iran

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2019.40477.1565
    http://ijmp.mums.ac.ir/article_13459.html
    https://iranjournals.nlai.ir/handle/123456789/324979

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب