      Effect of Phantom Size and Tube Voltage on the Size-Conversion Factor for Patient Dose Estimation in Computed Tomography Examinations

      (ندگان)پدیدآور
      Adhianto, DwiAnam, ChoirulSutanto, HeriAli, Mohd
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      1.309 مگابایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Technical Notes
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: This study aimed to establish the conversion factors to normalize the output dose of volumetric computed tomography dose index (CTDIvol) to the patient dose (i.e. size-specific dose estimate (SSDE)) for various phantom diameters and tube voltages. Material and Methods: In-house cylindrical acrylic phantoms with physical diameters ranging from 8 to 40 cm were developed in this study. Each phantom had a hole in the center and four holes in the peripheral areas. The phantoms were scanned by a Siemens Somatom Definition AS CT Scanner using different tube voltages (i.e. 80, 100, 120, and 140 kVps) and with 200 mAs and 10 mm slice thickness. In addition, the doses in every hole and phantom were measured using a Raysafe X2 CT Sensor. The weighted SSDE (SSDEw) values were calculated using the five holes in every measurement. The size-conversion factors for the body and the head CTDI phantoms were established by dividing the SSDEw for various sizes with the SSDEw at the water-equivalent diameter of 33.90 cm and 16.95 cm, respectively. Results: The results revealed that the size-conversion factor exponentially decreased with an increase in the phantom size. It was also found that the size-conversion factor was affected by the tube voltages. Furthermore, the different size-conversion factor between 80 and 140 kVp was more than 15% in very thin and obese patients. Conclusion: Higher accuracy of the size-specific dose estimation can be achieved considering the impact of the tube voltages beside the size of the patient.
      کلید واژگان
      ray Computed Tomography Patient Dose Radiation Dosimetry Radiation Dosage
      Computed Tomography (CT)
      Medical Physics
      Radiation Detection & Measurement

      شماره نشریه
      5
      تاریخ نشر
      2020-09-01
      1399-06-11
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Department of Physics, Faculty of Mathematics and Natural Sciences, Diponegoro University
      Department of Physics, Diponegoro University
      Department of Physics, Faculty of Mathematics and Natural Sciences, Diponegoro University
      Discipline of Medical Imaging Science, Faculty of Health Sciences, The University of Sydney

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2019.43059.1649
      http://ijmp.mums.ac.ir/article_14158.html
      https://iranjournals.nlai.ir/handle/123456789/324973

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب