• ورود به سامانه
      مشاهده مورد 
      •   صفحهٔ اصلی
      • نشریات انگلیسی
      • Iranian Journal of Medical Physics
      • Volume 17, Issue 4
      • مشاهده مورد
      •   صفحهٔ اصلی
      • نشریات انگلیسی
      • Iranian Journal of Medical Physics
      • Volume 17, Issue 4
      • مشاهده مورد
      JavaScript is disabled for your browser. Some features of this site may not work without it.

      An Enumeration Survey on Diagnostic X-Ray Generators and Essential Safety Parameters in Mizoram, India

      (ندگان)پدیدآور
      Tiwari, RameshLalrinmawia, JonathanPau, Kham
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      1.030 مگابایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: Best radiography practice involves operational optimal machine performance, delivering cost-effective healthcare services under appropriate safety conditions for workers and the public. The present study aimed to investigate the safety status of diagnostic X-ray installations in Mizoram, India. Material and Methods: Linearity of time (sec), linearity of current (mA), output reproducibility, table dose (μGy/mAs), peak voltage (kVp) accuracy, and 16 essential safety parameters of 135 X-ray machines were considered in this study. A battery-operated dosimeter and wide-range digital kVp meter were used to measure output radiation and effective peak potential of X-ray tube. Data analysis was performed using SPSS software to obtain the mean, standard deviation, and coefficient of variation. Results: Among different electronic parameters, 59.2% linearity of time, 82.6% linearity of current, 89.7% kVp accuracy, 35.1% output reproducibility, and 92.8% table dose were beyond the acceptable limits. Based on 16 essential safety parameters, it was observed that 98.7% of X-ray machines did not receive proper quality assurance test, 1.9% of the installations employed lead-line patient entrance doors, 46.8% of the machines were operated without any protective barriers and 83.1% of the units were operated without personnel monitoring service. Conclusion: The present study had concluded with more problems than the previous studies in different parts of the world in this regard. Due to the absence of proper quality control (QC) programs, many installations did not follow standard installation guidelines. The authors recommended that proper QC should be implemented by the frequent monitoring of each and every diagnostic X-ray installation.
      کلید واژگان
      Quality Assurance Diagnostic X
      Ray Radiation Protection
      Medical Physics

      شماره نشریه
      4
      تاریخ نشر
      2020-07-01
      1399-04-11
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Department of Physics, School of Physical Sciences, Mizoram University (A Central University), Tanhril campus, Aizawl
      Department of Physics, Mizoram University
      Mizoram State Cancer Institute, Zemabawk, Aizawl-796017, Mizoram, India

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2019.39307.1519
      http://ijmp.mums.ac.ir/article_13968.html
      https://iranjournals.nlai.ir/handle/123456789/324939

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب