      Assessment of Maximum Entrance Skin Dose of Patients Undergoing Cardiac Interventional Procedures and Its Correlation with Other Dosimetric Parameters

      (ندگان)پدیدآور
      Bahreyni Toossi, Mohammad TaghiKhosroabadi, MohsenMehrpouyan, MohammadRezaei Moghaddam, Homa
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      1.595 مگابایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: In recent years, the number of complex coronary angiography (CA) is increasing rapidly. These procedures have a significant contribution to medical exposure to the general population. Exposure of patients to high doses of x-rays could cause deterministic effects on the skin. Therefore, the assessment of radiation doses of patients is of great importance. This study aimed to assess maximum entrance skin dose (MESD) of patients who underwent interventional cardiology procedures. Moreover, it was attempted to determine the correlation between MESD and other relevant dosimetric parameters. Material and Methods: The MESDs of 32 patients who underwent CA procedures were measured by an array of thermoluminescence dosimeters (TLDs). In this study, a Perspex tray consisting of 5 rows and 6 columns was used to hold the TLDs. Its long axis was perpendicular to the long axis of the table, and the top edges of the tray were approximately equal to the patient's shoulders. Results: The results revealed a linear relationship between dose area product (DAP) values and MESDs (R2=0.89; P=0.00). In addition, there was a significant association between MESD and fluoroscopy time (R2=0.89). Moreover, a weak correlation was observed between MESD and the number of frames per second (R2=0.23). Conclusion: According to the results, the recorded DAP values and fluoroscopy time can be used to estimate the MESDs of patients undergoing coronary fluoroscopy procedures.
      کلید واژگان
      Coronary Angiography
      Fluoroscopy
      Radiation
      Dosimeter
      Medical Physics
      Radiation Detection & Measurement
      Radiation Protection (Ionizing & Non-Ionizing)

      شماره نشریه
      4
      تاریخ نشر
      2020-07-01
      1399-04-11
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Medical Physics Department, Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.
      Medical Physics Department, Faculty of Medicine, North Khorasan University of Medical Sciences, Bojnurd, Iran
      Radiology and Radiotherapy Department, Faculty of Medicine, Sabzevar University of Medical Sciences, Sabzevar, Iran
      Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2019.33109.1403
      http://ijmp.mums.ac.ir/article_13894.html
      https://iranjournals.nlai.ir/handle/123456789/324934

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب