      Measurements of Photon Beam Flattening Filter Using an Anisotropic Analytical Algorithm and Electron Beam Employing Electron Monte Carlo

      (ندگان)پدیدآور
      krabch, mohammed el adnaniChetaine, AbdelouahedSaidi, kamalERRADI, Fatima zohraNOURREDDINE, Abdelatibenkhouya, yassineEl Baydaoui, Redouane
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      1.191 مگابایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
        Introduction: This study aimed to report the measurement of photon and electron beams to configure the Analytical Anisotropic Algorithm and Electron Monte Carlo used in clinical treatment. Material and Methods: All measurements were performed in a large water phantom using a 3-dimensional scanning system (PTW, Germany). For photon beams, the data were measured with a 0.125cc cylindrical chamber. For electron, the data were performed with a Roos chamber. Results: In photon beams, flatness and symmetry for reference field size 10×10cm2 were within the tolerance intervals. Flatness were 0.79% and 1.55% for X6MV and X18MV, respectively. Symmetry were 0.57 and 0.25 for X6MV and X18MV, respectively. The output factor vary between 0.83 and 1.11 for X6MV. Moreover, it varies between 0.74 and 1.09 for X18MV. The leaf transmission factors were 0.97% for X6MV and1.14% for X18MV. The DLG were 1.31 and 1.34 for X6MV and X18MV, respectively. For electron beams, the quality index R50 for applicator 15×15cm2 were in the tolerance. Maximum depth dose for 6, 9, 12, 16 and 20MeV were 1.2, 1.9, 2.7, 2.99 and 2.4cm, respectively. Bremsstrahlung tail were 6MeV–2.86cm, 9MeV–4.32cm, 12MeV–5.96cm, 16MeV–7.93cm, and 20MeV–10.08cm per energy levels. Conclusion: The obtained results and international recommendations were in a good agrement
      کلید واژگان
      Linear accelerator Algorithm
      Monte Carlo Method Radiotherapy
      Medical Application of Computer Simulation
      Medical Physics
      Physics of Radiotherapy

      شماره نشریه
      3
      تاریخ نشر
      2019-05-01
      1398-02-11
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Nuclear reactor, nuclear security and environment group, Physics Department, Faculty of Sciences, Mohamed V University, Rabat, Morocco & Sheikh Khalifa Ibn Zaid Hospital
      Nuclear reactor, nuclear security and environment group, Physics Department, Faculty of Sciences, Mohamed V University, Rabat, Morocco
      Physics Department, Faculty of sciences, Mohamed V university, Rabat, Morocco
      Nuclear reactor, nuclear security and environment group, Physics Department, Faculty of Sciences, Mohamed V University, Rabat, Morocco
      Physics Department, Faculty of Science, Mohammed V University, Rabat, Morocco
      Physics Department,Faculty of sciences, Mohamed V university, Rabat, Morocco.
      Laboratory of technology and medical science, unity of biomedical instrumentation and medical physics, Higher Institute of Health Sciences; University Hassan I.

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2018.31544.1372
      http://ijmp.mums.ac.ir/article_11715.html
      https://iranjournals.nlai.ir/handle/123456789/324901

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب