    Field-In-Field Plan Versus Tangential Wedged Beam Plan in Chest Wall Radiotherapy of Post-Mastectomy Patients: Treatment Planning Study

    (ندگان)پدیدآور
    Yarahmadi, Mehranfaramarzi, baharhaghparast, abbassaalehi, zeinab
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    1.143 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: In this study, dose distribution of the chest wall in post-mastectomy breast cancer patients was evaluated and compared in the tangential wedged beam (TWB) and field-in-field (FIF) plans. Materials and Methods: Thirty-six patients with left-sided breast cancer were enrolled in this study. The FIF and TWB plans were generated for each patient to compare dosimetric parameters of the chest wall. The maximum dose (Dmax), homogeneity index (HI), conformity index (CI), and uniformity index (UI) were defined and used for comparison of the dosimetric parameters of the planning target volume (PTV) in both FIF and TWB plans. The percentage of volumes receiving at least 10, 20, 30, and 40 Gy of the left lung and 5, 10, 20, 25 and 30 Gy of the heart were used to compare the dosimetric results of the organs at risk. Statistical analysis was performed using SPSS, version 20. Results: The FIF plan had significantly lower HI (P=0.000) than the TWB plan, indicating that the FIF plan was better than the TWB plan in PTV. The V40lung (15.36±4.35 vs. 18.37±4.42) and V30heart (8.15±3.75 vs. 10.94±3.94; P=0.000) were significantly lower in the FIF plan than in the TWB plan. In addition, the monitor unit (MU) was significantly lower in the FIF plan than in the TWB plan (227.76 vs. 323.59; P=0.000). Conclusion: The FIF plan significantly reduced the dose volume of the left lung and heart in post-mastectomy radiotherapy compared to the TWB plan. Therefore, the FIF plan is recommended for this purpose.
    کلید واژگان
    Breast Cancer
    High-Energy X-Ray
    Field-In-Field Technique
    Treatment Planning
    TWB Plan
    Medical Physics
    Physics of Radiotherapy

    شماره نشریه
    4
    تاریخ نشر
    2017-12-01
    1396-09-10
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Medical Physics Dept., Kurdistan University of Medical Sciences, Sanandaj, Iran.
    Department of Medical Physics, School of medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran.
    Department of Medical Physics, Faculty of Medicine, Kermanshah, Iran.
    Tohid Hospital, Department of Radiotherapy, Kurdistan University of Medical Sciences, Sanandaj, Iran.

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2017.22982.1223
    http://ijmp.mums.ac.ir/article_9164.html
    https://iranjournals.nlai.ir/handle/123456789/324898

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب