    Determination of Optimum Planar Imaging Parameters for Small Structures with Diameters Less Than the Resolution of the Gamma Camera

    (ندگان)پدیدآور
    Nyathi, Mpumelelo
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    1.133 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: The limited spatial resolution of the gamma camera hinders the absolute quantification of planar images of small structures. The imaged structures are affected by partial volume effects (PVEs), which can spread activity and lead to underestimation of the regional distribution.  The use of optimum planar parameters reduces the impact of the limited spatial resolution of the gamma camera and the statistical noise inherent to low photon count, thus improving quantification. In this study, we aimed to determine the optimum planar imaging parameters for small structures. Materials and Methods: A thyroid protocol was used to acquire planar images of the spheres A, B, and C (16 mm, 12 mm, and 11 mm in diameter, respectively) whilst filled with a targeted activity concentration of technetium-99m. One sphere was mounted at the centre of the Jaszczak Phantom and the other two adjacent to its walls using capillary stems fitted on the spheres. The phantom was filled with distilled water. The targeted activity concentrations used were 74 kBq/mL, 100 kBq/mL, 150 kBq/mL, and 300 kBq/mL. Images of the same count per pixel were acquired on 64  64, 128  128, 256  256, 512  512, and 1024  1024 pixels using a vertical detector mounted 5 cm above the phantom. All the images were quantified using ImageJ software, version 1.48a, Java 1.70_51 [64-bit]. Results: The optimum planar imaging parameters established were a matrix size of 128 128 pixels and technetium-99m solution of activity concentration of 300 kBq/ml. Conclusion: The use of optimal imaging parameters reduces the impact of PVEs, leading to improved quantitative accuracy.
    کلید واژگان
    Medical Imaging
    Radioisotope Imaging
    Partial Volume Effects
    Medical Physics
    Physics of Nuclear Medicine

    شماره نشریه
    4
    تاریخ نشر
    2017-12-01
    1396-09-10
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Department of Medical Physics Sefako Makgatho Health Sciences University SouthAfrica

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2017.24559.1246
    http://ijmp.mums.ac.ir/article_9267.html
    https://iranjournals.nlai.ir/handle/123456789/324894

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب