      Evaluating the Effects of Field Size on Beam Homogeneity Coefficient in the Superficial Radiotherapy Machine Using Empirical Method and Simulation

      (ندگان)پدیدآور
      Babapour, HamedAllahverdi Pourfallah, TayyebShabestani Monfared, AliShirazi, Ali-Reza
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      417.1کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction Superficial X-ray therapy is one of the most important treatment methods in radiotherapy especially in the treatment of superficial skin lesions (Up to 300 kVp). Quality of the X-ray beam that can be expressed by Half-Value Layer (HVL), are important indices for this type of treatment effective energy of photon and Homogeneity Coefficient (HC). Materials and Methods The HC of the superficial X-ray machine at 180 kVp was determined by an experimental method and also by simulation (EGSnrc code) and the results were compared. The exposure after the first (0.5 mmCu) and second (1 mmCu) attenuation layers were measured by Farmer dosimeter for three different field sizes. The HC was derived from these experimental data and was compared with corresponding results acquired by EGSnrc code package. The BEAMnrc code was used to transport photons and electrons at exit level of the applicator, DOSXYZnrc code for calculation of absorbed dose in the phantom, and BEAMDP for drawing the output spectra. Results Results showed that the mean percentage difference of errors obtained by experimental data and simulation method is 3.3% and the average energy of output spectra in 180 kVp after passing through filters and attenuation layers increases to 15.3% to 25.4%. Conclusion The maximum percentage difference of errors for the applied methods is higher than analogous devolves elsewhere reports which could be due to setup parameters such as SSD and field size. However, the present method has the advantage of ease of setup and matching with clinical conditions.
      کلید واژگان
      EGSnrc
      Homogeneity Coefficient
      Ortho-Voltage Machine
      Radiotherapy
      Medical Application of Computer Simulation
      Medical Physics
      Physics of Radiotherapy

      شماره نشریه
      3
      تاریخ نشر
      2012-09-01
      1391-06-11
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Nuclear Engineering Department, Azad Islamic University- Sciences and Researches Unit, Tehran, Iran
      Biochemistry and Biophysics Department, Faculty of Medicine, Mazandaran University of Medical Sciences, Sari, Iran
      Biochemistry and Biophysics Department, Babol University of Medical Sciences, Shahid Rajaee Oncology Hospital, Babolsar, Iran
      Medical Physics Department, Tehran University of Medical Sciences, Tehran, Iran

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2012.142
      http://ijmp.mums.ac.ir/article_142.html
      https://iranjournals.nlai.ir/handle/123456789/324873

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب