      Investigating the Effects of Cut-Out Shield on High-Energy Electron Fields Using MAGIC Normoxic Polymer Gel

      (ندگان)پدیدآور
      Ansari Mehr1, HadisArbabi, AzimZahmatkesh, Mohammad HasanAllahverdi, MahmoodBagheri, Saeid
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      733.1کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction The use of cut-outs in electron applicators make changes on output, isodose, and percentage depth dose (PDD) curves. These changes and electron beam dose distribution in the form of three-dimensional (3D) can be measured by gel dosimeters. Materials and Methods Dosimetry was performed with and without a square shield (6×6 cm2 field). The energies were 4, 9, and 16 MeV and phantom was filled with MAGIC gel polymer. For each section, transverse relaxation rate (R2) maps were obtained from MRI images and percentage depth doses and isodose curves were plotted. Results Average energy was 3.029 MeV for the energy of 4 MeV and 8.155 MeV for the energy of 9 MeV. Surface dose was higher in shielded field compared with the open one (due to electron scattering between the phantom and lead) which increased with increasing of energy. In the open field, for energies equal to 4, 9, and 16 MeV, the surface dose was 6.40, 6.48, and 7.20 Gy and for the shielded mode, they were 6.63, 7.04, and 7.31 Gy, respectively. Also error values showed less errors and higher accuracy on curves by increasing of energy. Conclusion Investigation of an isodose pattern in the shielded mode showed scattering due to the lead, which is on the applicator. Overall, the results of this study demonstrated the value and potential of this dosimetric method with respect to characteristics such as stability, responsiveness and specially ability to show three-dimensional electron beam dose distribution.
      کلید واژگان
      Electron
      MAGIC polymer gel
      Shield
      Medical Physics
      Radiation Detection & Measurement

      شماره نشریه
      2
      تاریخ نشر
      2013-09-01
      1392-06-10
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Department of Medical Radiation Engineering, Science and Research Branch, Islamic Azad University, Tehran, Iran
      2- Department of Medical Physics, Imam Hosein Hospital, Shahid Beheshti Medical University, Tehran, Iran
      Department of Medical Physics, Novin Medical Radiation Institute, Tehran, Iran
      Department of Medical Physics, Tehran University of Medical Sciences, Tehran, Iran
      Novin Medical Radiation Institute, Shahid Beheshti University, Tehran, Iran

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2013.2175
      http://ijmp.mums.ac.ir/article_2175.html
      https://iranjournals.nlai.ir/handle/123456789/324809

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب