      A Comparative Study of the Construction of Positron Emission Tomography/Computed Tomography Facilities in Three South African Hospitals

      (ندگان)پدیدآور
      Masango, NompumeleloVan Wyk, BronwinMongane, Modisenyane
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      1.541 مگابایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: Development of higher energy modalities such as positron emission tomography/computed tomography (PET/CT), has led to more complex shielding problems. This is due to several factors, such as the radiopharmaceutical relatively high-administered activity, high patient throughput, and high energies of 511 kilo-electron volt (keV) positron annihilation photons. Therefore, this study aimed to compare three different methods used to determine the required shielding thicknesses of PET/CT facilities. Material and Methods: The required shielding thicknesses for three facilities were determined by using three different shielding methods, i.e. narrow beam, broad beam and Monte Carlo approximation. The design goal was chosen as 6 mSv/year for radiation workers and 1 mSv/year for the public. In addition, occupancy factors (T) were established, and all calculations had a use factor (U) of 1. The workload (W) of facilities and thicknesses of all barriers were then calculated for the three facilities. Results: For narrow beam approximation the average required thicknesses obtained were 6.16 mm lead, 5.12 cm concrete and 2.95 cm iron. Broad beam approximation required an average of 7.55 mm lead, 8.01 cm concrete and 2.96 cm iron thicknesses. Monte Carlo approximation required 7.62 mm lead, 10.59 cm concrete and 2.94 cm iron thicknesses. Conclusion: The narrow beam approximation demonstrated the least shielding thickness required for the materials used in this study, which can lead to under-shielding. The broad beam and Monte Carlo approximations demonstrated higher required shielding thickness although there were discrepancies between these two approximations for lead, concrete, and iron.
      کلید واژگان
      PET/CT
      Shielding
      Monte Carlo
      Medical Physics
      Physics of Nuclear Medicine

      شماره نشریه
      2
      تاریخ نشر
      2020-03-01
      1398-12-11
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Department of Medical Physics, Faculty of Health Science, Sefako Makgatho Health Science University, Pretoria, South Africa
      Department of Medical Physics, Faculty of Medicine, Dr George MUkhari Hospital, Pretoria, South Africa.
      University of free State Faculty: Health Sciences PO Box Bloemfontein 9300, Republic of South Africa

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2019.35911.1455
      http://ijmp.mums.ac.ir/article_13422.html
      https://iranjournals.nlai.ir/handle/123456789/324786

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب