      Estimation of Absorbed and Effective Doses in Organs through Computed Tomography Examinations Using Automatic Exposure Control and Fixed Tube Current Techniques: A Phantom Case Study

      (ندگان)پدیدآور
      Sulemana, HamzaInkoom, StephenSosu, EdemSchandorf, Cyril
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      1.274 مگابایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: The study aimed to assess absorbed and effective doses in organs through computed tomography (CT) examinations using automatic exposure control (AEC) and fixed tube current (FTC) techniques. Material and Methods: Scanning parameters were obtained for routine adult CT examinations and used to estimate the organ absorbed and effective doses using CT-Expo software. The estimated effective doses were based on International Commission on Radiological Protection publication 103 recommendations. Results: Regarding the scans performed with AEC, doses to head, chest, abdomen and pelvic organs were within the range of 19.7-41.8, 6.4-17.4, 19.2-20.9, and 10.5-24.9 mGy respectively. Moreover, the effective doses for the mentioned organs were 1.6, 6.1, 6.4 and 5.4 mSv respectively. Considering FTC technique, doses to organs ranged 16.7-75.5, 4.1-52.2, 10.6-33.2 and 5.2-38.7 mGy respectively. Moreover, the mean effective doses of FTC were 2.1, 6.9, 9.4 and 6.1 mSv, respectively. Examinations performed with AEC technique induced a dose reduction of 9% and 34% for head organs, 52, 62 and 25% for chest organs, 16% and 14% for abdomen organs, and 11% and 10% for pelvic organs, compared to the FTC. A dose increase of 3% was observed for testes. The mean effective doses for scans with AEC were 13-46% lower than those obtained by FTC. Conclusion:According to the obtained results of the current study, the estimated doses for scans with AEC technique were in a lower level compared to FTC technique. Accordingly, it is recommended to utilize this technique for CT examinations to ensure optimal dose reduction to radiosensitive organs.
      کلید واژگان
      Computed Tomography
      organ dose
      Radiation Dosimetry
      Radiology
      Computed Tomography (CT)
      Medical Physics
      Radiation Detection & Measurement

      شماره نشریه
      1
      تاریخ نشر
      2020-01-01
      1398-10-11
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Department of Medical Physics, University of Ghana, Accra, Ghana
      Radiation Protection Institute, Ghana Atomic Energy Commission and Department of Medical Physics, University of Ghana.
      Radio logical and Medical Sciences Research Institute, Ghana Atomic Energy Commission and Department of Medical Physics.
      Department of Medical Physics Graduate School of Nuclear and Allied Sciences, University of Ghana.

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2019.34196.1432
      http://ijmp.mums.ac.ir/article_12546.html
      https://iranjournals.nlai.ir/handle/123456789/324751

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب