      Effective Source-Surface Distance in Various Field Sizes and Electron Beam Energies and its Effect on Cutout Factor in a Elekta Precise Linear Accelerator

      (ندگان)پدیدآور
      Fallahi, FatemeBayatiani, Mohamad RezaSeif, FatemehAliasgharzadeh, AkbarMohaghegh, Fatholah
      Thumbnail
      نوع مدرک
      Text
      Conference Proceedings
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: In electron beam treatment, because of the non-point electron beam source, inverse-square law cannot be applied for dosimetry in different treatment intervals. Therefore, providing source-surface distance (SSD) charts in all clinics is of paramount importance. This study aimed to determine the effective SSD for various electron beam energies and field sizes and to evaluate its effect on cutout factor in a linear accelerator.   Materials and Methods: We used Elekta Precise linear accelerator in Ayatollah Khansari Hospital, Arak, Iran, for various energy levels (10, 15, and 18 MeV). The measurement environment was MP3-M water phantom (PTW Co., Canada), and diode detector was utilized for dosimetry. The effective SSD and cutout factor was estimated for 100, 105, 110, 115, and 120 cm SSDs and 1.5×1.5 and 20×20 cm2 square fields.   Results: The effective SSD in the 1.5×1.5 to 20×20 cm2 fields altered from 29.95 to 93.95 cm, 50.40 to 96.50 cm, and 63.51 to 95.32 cm for different energy levels of 10 MeV, 15 MeV, and 18 MeV, respectively. The cutout factor increased along with the field size, but decreased by extending the SSD. These alterations were more significant for the energy level of 10 MeV.   Conclusion: Since the effective SSD is dependent on energy level and field size, it is recommended to independently compute the effective SSD considering these variables. Furthermore, for designing accurate therapies, cutout factor variations should be considered for small-sized fields, especially at low energy levels.
      کلید واژگان
      Cutout factor
      Electron Beams
      Effective SSD
      Field

      شماره نشریه
      12
      تاریخ نشر
      2018-12-01
      1397-09-10
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Department of Medical Physics, School of Paramedical Sciences, Kashan University of Medical Sciences, Kashan, Iran
      Department of Medical Physics and Rariotherapy, Arak University of Medical Sciences and Khansari Hospital, Arak, Iran
      Department of Medical Physics and Rariotherapy, Arak University of Medical Sciences and Khansari Hospital, Arak, Iran
      Department of Medical Physics, School of Paramedical Sciences, Kashan University of Medical Sciences, Kashan, Iran
      Department of Medical Physics and Rariotherapy, Arak University of Medical Sciences and Khansari Hospital, Arak, Iran

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2018.13035
      http://ijmp.mums.ac.ir/article_13035.html
      https://iranjournals.nlai.ir/handle/123456789/324657

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب