    Evaluation of the gray level in CBCT systems and its relationship with HU in CT Scanners

    (ندگان)پدیدآور
    Aghaz, AhdiyehKardan, Mohammad RezaDeevband, Mohammad RezaBahadorzadeh, BahadorKasesaz, YaserGhadiri, Hossein
    Thumbnail
    نوع مدرک
    Text
    Conference Proceedings
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: Cone-beam CT (CBCT) is an imaging system which offers three-dimensional (3D), multiplanar images and has many advantages over computed tomography (CT) including shorter acquisition times for the resolution desired in dentistry, lower radiation dose to the patient, reasonable price and higher spatial resolution but CBCT scanners are unable to display actual Hounsfield units (HU) similar to medical CT. In CT scan, to show the image that represents the density of the tissue Hounsfield Unit is proportional to the degree of x-ray attenuation and it is allocated to each pixel. In CBCT systems displayed gray levels are arbitrary and so variable, so the derived density provided less than meaningful data and the ability to assess the density or quality of bone is limited. The purpose of the present study was to present a clinical study demonstrating a method to derive Hounsfield units from gray levels in cone beam CT (CBCT). Materials and Methods: For the evaluation of CBCT gray values, a new versatile quality control tool of dental Cone- Beam CT, Pro-Dent CT mk II phantom, which consists of a main PMMA cylinder that houses modules with different test objects was used. Four CBCT devices were used for the scanning of the phantom: CS8100, CS9300(Carestream Health, New York, NY), New Tom GIANO (Quantitative Radiology, Verona, Italy), Vatech Pax-i3D (Hwaseong-si, Gyeonggi-do, Korea). The reconstructed data were exported as Digital Imaging and Communications in Medicine (DICOM) and analyzed with On Demand 3DH by Cybermed, Seoul, Korea and ImageJ as an image processing toolkit. The relationship between gray levels and linear attenuation coefficients in various kV and mA was investigated. Results: The results indicated a strong linear relationship between the gray scales in CBCT and HU in CT using the standard definition HU = (µmaterial   –  µwater)/(µwater) × 1000.  This made it possible to calculate Hounsfield units from the measured gray levels. Uncertainty in determining effective energies for each of CBCT systems resulted in unrealistic effective energies and significant variability of calculated CT numbers. Linear regression (0.802 Conclusion: Considering the fact that CBCT gray levels are inaccurate to rely upon for decisions on implant placement. HU can be derived from the gray levels in dental CBCT scanners using linear attenuation coefficients as an intermediate step. This study presented a method to convert gray levels in CBCT imaging into meaningful HU. Establishment of meaningful HU for CBCT will open new possibilities for clinicians in implant planning, diagnosis, surgical interventions, treatment planning, 2D and 3D reconstruction of images.
    کلید واژگان
    Hounsfield units
    Cone beam computed tomography
    Gray levels
    Bone Density
    Linear attenuation coefficient

    شماره نشریه
    12
    تاریخ نشر
    2018-12-01
    1397-09-10
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Nuclear Science and Technology Research Institute, Tehran, Iran, mkardan@aeoi.org.ir.
    Nuclear Science and Technology Research Institute, Tehran, Iran,
    Department of Medical Physics, School of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
    Nuclear Engineering Department, Shiraz University, Shiraz, Iran. Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran.
    Nuclear Science and Technology Research Institute, Tehran, Iran,
    Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran.

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2018.12847
    http://ijmp.mums.ac.ir/article_12847.html
    https://iranjournals.nlai.ir/handle/123456789/324515

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب