      Evaluation of dose distribution and dose gradient in brachytherapy cylindrical applicators using a dedicated Phantom for Iridium-192 and Cobalt-60 HDR sources

      (ندگان)پدیدآور
      Gholami, Mohammad HadiMohammadi, MohammadSedighi, AbdolazimJaberi, RaminRezaei, Fahimeh
      Thumbnail
      نوع مدرک
      Text
      Conference Proceedings
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: A study was performed to evaluate radiation dose distribution and dose gradient around cylindrical applicators for high-dose-rate (HDR) brachytherapy systems with 192Ir, 60Co brachytherapy source applied for rectal and vaginal cancers treatments.   Materials and Methods: Two additional computed tomography (CT) based plans were generated using a dedicated Phantom for both 60Co source and 192IridumHDR sources. The phantom contains a film and TLD-100 position on the top and peripheral areas, and a rectum cylinder with 1.5 cm insert caps into the phantom. Dose coverage of 5-mm shell surrounding the cylinder, was then evaluated for both scenarios. Results:A significant dose reduction from the expectation dose for 192Ir at the central points of the cylinder tip was observed. The results were then verified by measurements through radio- chromic film and TLDs.   Conclusion: Both 60Co and 192Ir sources were found to be suitable HDR sources for vaginal treatment using a cylindrical applicator with advantages of reducing radiation exposure to Organ At Risks (OARs) in the lower dose range, and appropriate prescribed dose to the target
      کلید واژگان
      Dose Distribution
      Cylinder Applicators
      Brachytherapy
      High Dose Rate Brachytherapy

      شماره نشریه
      12
      تاریخ نشر
      2018-12-01
      1397-09-10
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Islamic Azad University, Science and research branch, Tehran, Iran Mahdieh Radiotherapy and Oncology charity center, Hamadan, Iran
      Department of Medical Physics, Royal Adelaide Hospital, Adelaide, SA 5000 AUSTRALIA. School of Physical Sciences, the University of Adelaide, SA 5005 AUSTRALIA. Department of Medical Physics, Reza Radiotherapy and Oncology Center, Mashhad, Iran
      Department of radiation oncology, University of medical science, Hamedan, Iran
      Department of Radiation Oncology, Tehran University of Medical Sciences, Tehran, Iran
      MSc of Medical physics, Department of Medical physics, Kermanshah University of Medical sciences, Kermanshah, Iran

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2018.11879
      http://ijmp.mums.ac.ir/article_11879.html
      https://iranjournals.nlai.ir/handle/123456789/324317

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب