      Comparison of the Obtained Results from Two Calibration Methods in MRI Brachytherapy Normoxic Polymer Gel Dosimetry

      (ندگان)پدیدآور
      Sedaghat, M.Zahmatkesh, M.HJaberi, R.Akhalghpoor, Sh.Allahverdi, M.
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      782.8کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: Despite the fact that the clinical implementation of polymer gel dosimetry has been facilitated after the introduction of normoxic gels, there are still complications in its clinical routine use that are under investigation. In the current work, the feasibility of using a normoxic polymer gel dosimeter named MAGICA has been investigated for use in our clinical brachytherapy applications at ImamHospital and the deviations between the results obtained from two different calibration methods were investigated. Materials and Methods: The gel fabrication was performed at Novin Medical Radiation Institute and 4 Perspex walled phantoms designed especially for brachytherapy irradiations were filled with the same gel, together with 15 plastic calibration test tubes. The test tubes were irradiated with a range of known doses using a cobalt-60 teletherapy unit and the phantoms were irradiated with two brachytherapy remote-after-loaders using Cs137 and Co60 brachytherapy sources according to the treatment planning data. Imaging was performed with a Multi-spin-echo protocol using a Siemens 1.5T MRI machine. Image processing was performed in MATLAB environment to extract R2 maps of the dose distributions. Gel fabrication, irradiation and imaging were performed at three different centers, so the dosimeters and test tubes were transferred from one center to another under non-uniform conditions. R2 values of the first dosimeter which was irradiated with a single cesium source were extracted from different distances from the source along a horizontal profile and correlated to the dose values at the same distances in treatment planning data. This was used as the second calibration method to extract the absolute measured values of the other three dosimeters. Diagrams of the absorbed dose versus distance were plotted for each dosimeter and the results obtained from the two calibration methods were compared. Results: Results of the measured values that were calibrated with the test tubes show a 3 to 6 Gy dose difference in plateaus and more than 2 mm distance to agreement gap in steep dose gradients. However, the results of using the second calibration method show at most a 0.36 Gy dose difference in plateaus and less than 2 mm distance to agreement gap in steep dose gradients. Discussion and Conclusion: As the gel fabrication, irradiation and imaging conditions were the same for all the gel dosimeters and the calibration tubes, the unexpected deviations between the results of the two calibration methods deserves more investigations. It seems plausible to attribute this deviation to the difference in the size of calibration test tubes and original dosimeters. From a practical point of view, as the origin of this effect has not yet been investigated, it would be best to use a calibration phantom with the same size and characteristics as the original dosimeters.
      کلید واژگان
      Polymer Gel Dosimetry
      Normoxic
      Brachytherpy
      MRI
      Calibration
      MAGICA
      Magnetic Resonance Imaging (MRI)
      Medical Physics
      Physics of Radiotherapy
      Radiation Detection & Measurement

      شماره نشریه
      1
      تاریخ نشر
      2006-03-01
      1384-12-10
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Ph.D. Student in Medical Physics, University of Sherbrooke, Canada.
      Assistant Professor, Novin Medical Radiation Institute, Tehran, Iran.
      M.Sc. in Medical Physics, Cancer Institute, Imam Hospital, TUMS, Tehran, Iran.
      Assistant Professor, Faculty of Medicine, TUMS & Novin Medical Radiation Institute, Tehran, Iran.
      Assistant Professor, Dept. of Medical Physics, Faculty of Medicine, TUMS & Cancer Institute of Imam Hospital, Tehran, Iran.

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2006.8127
      http://ijmp.mums.ac.ir/article_8127.html
      https://iranjournals.nlai.ir/handle/123456789/324255

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب