      Monte Carlo Modeling of Dual and Triple Photon Energy Absorptiometry Technique

      (ندگان)پدیدآور
      Kamali-Asl, AlirezaDarafsheh, ArashAghamiri, Seyed Mahmoudreza
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      423.1کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: Osteoporosis is a bone disease in which there is a reduction in the amount of bone mineral content leading to an increase in the risk of bone fractures. The affected individuals not only have to go through lots of pain and suffering but this disease also results in high economic costs to the society due to a large number of fractures.  A timely and accurate diagnosis of this disease makes it possible to start a treatment and thus preventing bone fractures as a result of osteoporosis. Radiographic methods are particularly well suited for in vivo determination of bone mineral density (BMD) due to the relatively high x-ray absorption properties of bone mineral compared to other tissues. Materials and Methods: Monte Carlo simulation has been conducted to explore the possibilities of triple photon energy absorptiometry (TPA) in the measurement of bone mineral content. The purpose of this technique is to correctly measure the bone mineral density in the presence of fatty and soft tissues. The same simulations have been done for a dual photon energy absorptiometry (DPA) system and an extended DPA system. Results: Using DPA with three components improves the accuracy of the obtained result while the simulation results show that TPA system is not accurate enough to be considered as an adequate method for the measurement of bone mineral density. Discussion: The reason for the improvement in the accuracy is the consideration of fatty tissue in TPA method while having attenuation coefficient as a function of energy makes TPA an inadequate method. Conclusion: Using TPA method is not a perfect solution to overcome the problem of non uniformity in the distribution of fatty tissue.
      کلید واژگان
      Bone mineral density (BMD)
      Dual photon energy absorptiometry (DPA)
      Triple photon energy absorptiometry (TPA)
      Monte Carlo modelling
      MCNP
      Medical Application of Computer Simulation
      Medical Imaging & Image Processing
      Medical Physics

      شماره نشریه
      34
      تاریخ نشر
      2007-12-01
      1386-09-10
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      Assistant Professor, Radiation Medicine Dept., Shahid-Beheshti University, Tehran, Iran
      M.Sc. in Radiation Medicine, Shahid-Beheshti University, Tehran, Iran
      Associate Professor, Radiation Medicine Dept., Shahid-Beheshti University, Tehran, Iran

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2007.7552
      http://ijmp.mums.ac.ir/article_7552.html
      https://iranjournals.nlai.ir/handle/123456789/324236