      Verification of the Accuracy of the Delivered Dose in Pelvic and Breast Cancer Radiotherapy by in-vivo Semi-Conductor Dosimetry

      (ندگان)پدیدآور
      Shabani, HamidrezaAllahverdi, MahmoodEsfahani, MahbodAghili, MehdiTakavar, AbbasTaghizade, Mohammd RezaTizmagh, ZahraVaezzade, Ali
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      1.172 مگابایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Paper
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: Delivering maximum dose to tumor and minimum dose to normal tissues is the most important goal in radiotherapy. According to ICRU, the maximum acceptable uncertainty in the delivered dose compared to the prescribed dose should be lower than 5%, and this is because of the relationship between absorbed dose, tumor control and normal tissue damage. Absorbed dose accuracy is investigated by an in vivo dosimetry method. In this paper, we compared absorbed dose in the tumors of the breast and pelvic region against the calculated dose. The amount of deviations and the factors that cause this deviation in dose delivery to patients and some methods for decreasing them were evaluated. Materials and methods: The entrance and exit doses of 36 pelvic-region cancer patients and 38 breast cancer patients who were treated by cobalt-60 teletherapy were measured using p-type diodes. It should be noted that the transmission method was used to assess the dose at isocenter. Two ionization chambers (0.6 cc and 0.3 cc) were used for calibration and determination of the correction coefficients in water and slab phantoms. Deviations between calculated and measured doses of entrance, exit and midline points were calculated and the results were shown using histograms. Results: The average and standard deviation for entrance, exit and midline points for pelvis cancer were assessed to be about 0.10%, -1.86% and -1.35% for mean deviation and 5.03%, 7.32% and 5.86% for standard deviation, respectively. The corresponding data for breast cancer were 0.78%, 5.29% and 3.59% for mean deviation and 5.97%, 10.23% and 9.86%, respectively. There was no significant difference between the calculated and measured doses (p > 0.1), except exit dose in breast cancer (p 0.1), except exit dose in breast cancer (p Discussion and Conclusions: Some error sources are patient setup error, patient motion and dose calculation algorithm error (due to ignoring inhomogeneity and patient curvature). As no significant deviations were found in midline dose, the method used has an acceptable accuracy. In vivo dosimetry can perform a basic role in the quality control of radiotherapy departments.
      کلید واژگان
      Radiotherapy of Breast and Pelvis
      In Vivo Dosimetry
      Semi-Conductor Dosimeter
      Ionization Chamber
      Medical Physics
      Physics of Radiotherapy
      Radiation Detection & Measurement

      شماره نشریه
      34
      تاریخ نشر
      2009-12-01
      1388-09-10
      ناشر
      Mashhad University of Medical Sciences
      سازمان پدید آورنده
      M.Sc. Student of Medical Physics, Department of Medical Physics, Tehran University of Medical Sciences, Tehran, Iran
      Associated Professor, Medical Physics Dept., Tehran University of Medical Sciences, and Radiotherapy Physics Dept., of the Cancer Institute, Tehran, Iran
      Physisist of the Radiotherapy Physics Department of the Cancer Institute, Tehran, Iran
      Associate Professor, Radiotherapy and Oncology Department of the Cancer Institute, Imam Khomeini Hospital, Tehran, Iran
      Professor, Medical Physics Dept., Tehran University of Medical Sciences, Tehran, Iran
      M.Sc. of Medical Physics, Imam Khomeini Medical Imaging Center, Gilan University of Medical Sciences, Some'e-sara, Iran
      Physisist of the Radiotherapy Physics Department of the Cancer Institute, Tehran, Iran
      M.Sc. Student of Medical Physics, Department of Medical Physics, Tehran University of Medical Sciences, Tehran, Iran

      شاپا
      2345-3672
      URI
      https://dx.doi.org/10.22038/ijmp.2009.7287
      http://ijmp.mums.ac.ir/article_7287.html
      https://iranjournals.nlai.ir/handle/123456789/324200

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب