    Evaluation of Dose Distribution Accuracy in HDR Brachytherapy of Esophagus Cancer Based on MRI Normoxic Polymer Gel Dosimetry

    (ندگان)پدیدآور
    Bahreyni Toossi, Mohammad TaghiSafaeian, GhorbanBahreyni Toossi, Mohammad HoseinBaiani, Shahram
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    995.7کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: The purpose of this work was to study the ability of MRI normoxic polymer gel dosimetry for evaluating the dose distribution in HDR brachytherapy of esophagial cancer at Imam Reza brachytherapy center (Mashhad, Iran). Materials and Methods: Initially, 2liters of normoxic gel (MAGIC) was fabricated and then poured into 12 calibration test tubes and placed in a perspex walled phantom. The gel phantom was irradiated with a brachytherapy remote-afterloader unit using a cobalt-60 brachytherapy source and the test tubes were irradiated with a range of known doses with a cobalt-60 teletherapy unit. Imaging was performed with a multi-spin-echo protocol and a T2 quantitative technique using a Siemens 1.5 T MRI machine. The MRI images were transferred to a computer and then image processing was performed in the MATLAB environment to extract R2 maps of the irradiated area. Results: In this study and at the reference point, the dose deviation between the gel dosimetry and the calculated data was 4.5%. The distance to agreement (DTA) for dose profiles was 2.7 mm. Also, dose sensitivity of the MAGIC gel dosimeter was 0.693 S-1Gy-1 (R2 =0.9376). Conclusion: In this work, the data obtained from TPS calculations were found in very good agreement with the measured results provided by gel dosimetry. It was evaluated using a comparison of isodoses and dose at the reference point, and dose profile verification. It is also concluded that the gel dosimetry systems have proven to be a useful tool for dosimetry in clinical radiotherapy applications.
    کلید واژگان
    MRI gel dosimetry
    gel calibration
    MAGIC
    TLD
    Brachytherapy
    Medical Physics
    Physics of Radiotherapy
    Radiation Detection & Measurement

    شماره نشریه
    1
    تاریخ نشر
    2010-03-01
    1388-12-10
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Professor, Medical Physics Research Center, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran
    M.Sc., in Medical Physics, Mashhad University of Medical Sciences, Mashhad, Iran
    Professor, Medical Physics Research Center, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran
    M.Sc., in Medical Engineering, Medical Physics Research Center, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2010.7267
    http://ijmp.mums.ac.ir/article_7267.html
    https://iranjournals.nlai.ir/handle/123456789/324188

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب