      Evaluation of the performance of parallel-hole collimator for high resolution small animal SPECT: A Monte Carlo study

      (ندگان)پدیدآور
      Sadremomtaz, AlirezaTelikani, Zeinab
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      760.8کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Article
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: Image quality and accuracy of in vivo activity quantification in SPECT are affected by collimator penetration and scatter components, especially in high energy imaging. These phenomena highly depend on the collimator characteristic and photon energy. The presence of penetrated and scattered photons from collimator in SPECT images degrades spatial resolution, contrast and image quality. Knowledge of penetration and scatter distribution is essential for optimization of collimator design and development of reconstruction algorithms.The aim of this study to survey the collimator performance of the newly developed HiReSPECT dual head gamma camera with pixelated array CsI(Na). Methods:We modeled the HiReSPECT, by using SIMIND Monte Carlo simulation code. The contribution of geometric, scatter and penetration components were quantitatively calculated for the different energy sources. Then we compared these results with simulation results of another small animal SPECT with compact pixelated array CsI(Tl) detector. Results:The simulated System spatial resolution and energy resolution of the HiReSPECT at 140keV respectively are 1.9mm and 29.72 keV (21.23%) FWHM  at 2.5cm distance from detector surface also Geometric, penetration, and scatter at 140keV for the HiReSPECT collimator are 96.42%, 2.22%, 1.30%, respectively. Similarly, geometric, penetration, and scatter at 159keV and 245keV for this system collimator are (95.24%, 3.08%, 1.68%) and (87.21%, 8.10%, 4.69%), respectively. Conclusion: The results verified that the magnitude of these components depend on collimator geometric structure and photons energy. The measured performances indicated that the HiReSPECT scanner is well suited for preclinical molecular imaging research and provide high resolution for small animal imaging.
      کلید واژگان
      SPECT
      Compact pixelated gamma camera
      Septal penetration
      Scatter
      Monte Carlo Simulation
      Medical Physics, Dosimetry, Instrumentation and Radiobiology

      شماره نشریه
      2
      تاریخ نشر
      2016-07-01
      1395-04-11
      ناشر
      Research Center for Nuclear Medicine (Tehran University of Medical Sciences)
      سازمان پدید آورنده
      Department of Physics, Faculty of Science, University of Guilan, Rasht, Iran
      Department of Physics, Faculty of Science, University of Guilan, Rasht, Iran

      شاپا
      1681-2824
      2008-2509
      URI
      http://irjnm.tums.ac.ir/article_20597.html
      https://iranjournals.nlai.ir/handle/123456789/300504

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب