    The influence of using different reconstruction algorithms on sensitivity of quantitative 18F-FDG-PET volumetric measures to background activity variation

    (ندگان)پدیدآور
    Ketabi, AliGhafarian, PardisMosleh-Shirazi, MohammadMahdavi, SeyedAy, Mohammad Reza
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    1.296 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Article
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: This study aims to investigate the influence of background activity variation on image quantification in differently reconstructed PET/CT images. Methods: Measurements were performed on a Discovery-690 PET/CT scanner using a custom-built NEMA-like phantom. A background activity level of 5.3 and 2.6 kBq/ml 18F-FDG were applied. Images were reconstructed employing four different reconstruction algorithms: HD (OSEM with no PSF or TOF), PSF only, TOF only, and TOFPSF, with Gaussian filters of 3 and 6.4 mm in FWHM. SUVmax and SUVpeak were obtained and used as cut-off thresholding; Metabolic Tumor Volume (MTV) and Total Lesion Glycolysis (TLG) were measured. The volume recovery coefficients (VRCs), the relative percent error (ΔMTV), and Dice similarity coefficient were assessed with respect to true values. Results:SUVmax and SUVpeak decreased and MTV increased as function of increasing the background dose. The most differences occur in smaller volumes with 3-mm filter; Non-TOF and Non-PSF reconstruction methods were more sensitive to increasing the background activity in the smaller and larger volumes, respectively. The TLG values were affected in the small lesions (decrease up to 12%). In a range of target volumes, differences between the mean ΔMTV in the high and low background dose varied from -11.8% to 7.2% using SUVmax and from 2.1% to 7.6% using SUVpeak inter reconstruction methods. Conclusion: The effect of the background activity variation on SUV-based quantification in small lesion was more noticeable than large lesion. The HD and TOFPSF algorithms had the lowest and the highest sensitivity to background activity, respectively.
    کلید واژگان
    18F-FDG-PET/CT
    Background activity
    Image reconstruction
    MTV
    TLG
    Medical Physics, Dosimetry, Instrumentation and Radiobiology

    شماره نشریه
    2
    تاریخ نشر
    2018-07-01
    1397-04-10
    ناشر
    Research Center for Nuclear Medicine (Tehran University of Medical Sciences)
    سازمان پدید آورنده
    Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran|Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran
    Chronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases, Shahid Beheshti University of Medical Sciences, Tehran, Iran|PET/CT and Cyclotron Center, Masih Daneshvari Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran
    Medical Imaging Research Center and Physics Unit, Department of Radiotherapy and Oncology, Namazi Hospital, Shiraz University of Medical Sciences, Shiraz, Iran
    Department of Medical Physics, Faculty of Medicine, Iran University of Medical Sciences, Tehran, Iran
    Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran|Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran

    شاپا
    1681-2824
    2008-2509
    URI
    http://irjnm.tums.ac.ir/article_31623.html
    https://iranjournals.nlai.ir/handle/123456789/300491

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب