CHARACTERIZATION OF A NEW PREPARED Li2B4O7: Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY
(ندگان)پدیدآور
SHARIFZADEH, M.GORJIFARD, R.نوع مدرک
TextResearch Note
زبان مدرک
Englishچکیده
A lithium borate thermoluminescence dosimeter with 0.3 wt% of manganese impurityconcentration was prepared for high dose levels in the range of 1-30 kGy for gamma irradiation applications.The TLD reader was used for read out purposes from ambient temperature to about 400°C with a heating rate of 6°C/sec. The pre- and post- irradiation anneal temperature was 300°C for 1.5 and 1 hours respectively. The effects of absorbed dose, dose rate, storage time, as well as the temperature and humidity versus pre- and post-irradiation storage time have been studied. The main glow peak was found to occur at 375°C, and the dosimeter response in the range of 1-30 kGy was linear. No significant fading effect of the dosimetric peak was observed after 8 weeks of storage, and the effects of ambient temperature and humidity were less than ±5% at 95% confidence level.
کلید واژگان
DosimetryThermoluminescence
dose response
dose rate
glow curve
شماره نشریه
3تاریخ نشر
2008-09-011387-06-11
ناشر
Springerسازمان پدید آورنده
Radiation Applications Research School, Nuclear Science &Technology Research Institute, P.O. Box 11365-3486, Tehran, I. R. of IranRadiation Applications Research School, Nuclear Science &Technology Research Institute, P.O. Box 11365-3486, Tehran, I. R. of Iran