    Validation of the Gel & Wax Boluses and Comparison of their Dosimetric Performance with Virtual Bolus

    (ندگان)پدیدآور
    Verma, T RPainuly, N KTyagi, MJohny, DGupta, RBhatt, M L B
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    850.2کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Research
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Background: In general, radiotherapy treatment planning is performed using the virtual bolus. It is necessary to investigate physical bolus in comparison to virtual one. Objectives: In the present study, first, radiological properties of superflab Gel bolus and Paraffin wax bolus was investigated in terms of their relative electron density. Then, dosimetric performance of both the bolus (i.e. Gel and Parafin wax) was compared with Virtual bolus.Material and Methods: In This experimental study, the radiological property of Wax and Gel boluses was investigated using two methods. In one, the relative electron density of both the Gel and Wax boluses was calculated by measuring their linear attenuation coefficient where in another method relative electron density was calculated by recording their CT No directly from their CT scan. Later CT scan of solid water slab phantom (dimension 30x30x15 cm3), with physical boluses (i.e. Gel and Wax bolus) of appropriate thicknesses required to deliver a dose of 200 cGy at Dmax using 4 MV, 6 MV and 15 MV photon beams, was taken. These CT data sets were retrieved to TPS. A plan was done to deliver a dose of 200 cGy at Dmax using Single 4 MV, 6 MV and 15 MV photon beams. Dose at depths Dmax, 1 cm, 2 cm, 3 cm, 4 cm and 5 cm was recorded. Using this similar method, doses at depths viz Dmax, 1 cm, 2 cm, 3 cm, 4 cm and 5 cm was recorded for the Gel and Wax boluses. The differences in dose of gel and wax bolus from virtual bolus were recorded for comparison of their dosimetric performance. Results: The measured (calculated) relative electron density of wax and Gel bolus was found to be 0.958 (0.926) and 0.923 (0.907), respectively. Variation in dosimetric performance of Gel and Wax with reference to Virtual bolus was studied. However, on average, Gel bolus was more consistent with virtual bolus. Conclusion: To avoid any dose difference between, delivered (using physical bolus) and planned (using virtual bolus), the physical boluses should be investigated for their dosimetric performance in comparison to virtual bolus. The results obtained and methodology used in this study can be applied in routine radiotherapy practices.
    کلید واژگان
    Paraffin
    Tomography, X-Ray Computed
    Imaging
    Waxes

    شماره نشریه
    6
    تاریخ نشر
    2019-12-01
    1398-09-10
    ناشر
    Shiraz University of Medical Sciences
    سازمان پدید آورنده
    PhD, Department of Radiotherapy, King George's Medical University, UP, Lucknow, India
    PhD, Department of Radiotherapy, King George's Medical University, UP, Lucknow, India
    MSc, Department of Radiology & Radiotherapy, UCMS, Delhi, India
    MSc, Department of Radiotherapy, King George's Medical University, UP, Lucknow, India
    MD, Department of Radiotherapy, King George's Medical University, UP, Lucknow, India
    MD, Department of Radiotherapy, King George's Medical University, UP, Lucknow, India

    شاپا
    2251-7200
    URI
    https://dx.doi.org/10.31661/jbpe.v0i0.1177
    https://jbpe.sums.ac.ir/article_45886.html
    https://iranjournals.nlai.ir/handle/123456789/26546

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب