    An Empirical Transmitted EPID Dosimetry Method using a Back-Projection Algorithm

    (ندگان)پدیدآور
    Hashemi, S MBahreyni, M HMohammadi, MNasseri, SBayani, SGholamhosseinian, HSalek, RShahedi, FMomennezhad, M
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    817.0کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Research
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Background: The present study aimed to introduce a rapid transmission dosimetry through an electronic portal-imaging device (EPID) to achieve two-dimensional (2D) dose distribution for homogenous environments. Material and Methods: In this Phantom study, first, the EPID calibration curve and correction coefficients for field size were obtained from EPID and ionization chamber. Second, the EPID off-axis pixel response was measured, and the grey-scale image of the EPID was converted into portal dose image using the calibration curve. Next, the scattering contribution was calculated to obtain the primary dose. Then, by means of a verified back-projection algorithm and the Scatter-to-Primary dose ratio, a 2D dose distribution at the mid-plane was obtained. Results: The results obtained from comparing the transmitted EPID dosimetry to the calculated dose, using commercial treatment planning system with gamma function while there is 3% dose difference and 3mm distance to agreement criteria, were in a good agreement. In addition, the pass rates of γ Conclusion: Based on the results, the method proposed can be used in EPID dosimetry.
    کلید واژگان
    Radiotherapy Planning, Computer-Assisted
    Dose-Response Relationship
    Radiotherapy
    Algorithms

    شماره نشریه
    5
    تاریخ نشر
    2019-10-01
    1398-07-09
    ناشر
    Shiraz University of Medical Sciences
    سازمان پدید آورنده
    PhD, Medical Physics Research Center, Mashhad University of medical Science, Mashhad, Iran
    PhD, Medical Physics Research Center, Mashhad University of medical Science, Mashhad, Iran
    PhD, Department of Medical Physics, Royal Adelaide Hospital, Adelaide, SA 5000, Australia
    PhD, Medical Physics Research Center, Mashhad University of medical Science, Mashhad, Iran
    MSc, Medical Physics Research Center, Mashhad University of medical Science, Mashhad, Iran
    PhD, Medical Physics Research Center, Mashhad University of medical Science, Mashhad, Iran
    MD, Cancer Research Center, Mashhad University of medical Science, Mashhad, Iran
    MSc, Department of Radiology, School of Paramedical Sciences, Torbat Heydariyeh University of Medical Sciences, Torbat Heydariyeh, Iran
    PhD, Medical Physics Research Center, Mashhad University of medical Science, Mashhad, Iran

    شاپا
    2251-7200
    URI
    https://dx.doi.org/10.31661/jbpe.v0i0.1082
    https://jbpe.sums.ac.ir/article_45725.html
    https://iranjournals.nlai.ir/handle/123456789/26536

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب