    Personalized 18FDG Dose Synthesis Using BG-75 Generator: 1st Year Experience at JCI Accredited Tertiary Care Hospital in Pakistan

    (ندگان)پدیدآور
    Zaman, M UFatima, NMehdi, S USami, MKhan, KhKhan, G
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    642.9کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Research
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Background: Compact cyclotrons are getting popular to fulfill enormous current demands of PET tracers. Aga Khan University Hospital, Karachi, Pakistan has acquired the first smallest footprint of BG-75 Generator for 18FDG-based PET/CT clinical imaging. We are sharing our experience of BG-75 in the first year (December 2015-November 2016) after commissioning. Material and Methods: BG-75 Generator (ABT, USA) was installed in available space without major design modification. It has a self-shielded mini-cyclotron (7.5 MeV proton beam, positive ion with current Results: Total 545 runs in 167 days (3±01 runs/day) were made. Average yield with 60 minute bombardment using 4.5 µA current was 37 mCi and 21 mCi for [18F]F- and 18FDG, respectively. Total 29 runs had chemistry or QC failures and were discarded. Remaining 516 batches were used to perform imaging upon 1370 patients (8±03 patients/day). Radiation dose in BG-75 suite and effective doses to 02 operators were within statutory limits. Conclusion: BG-75 Generator provides safe, dependable and sustainable supply of 18F-Fluoride for 18FDG and other low-volume clinical PET imaging. Its compactness and automation need minimal space and manpower. Radiation dose rate in cyclotron suite and personal dosimetry were also found within safe limits. Its Dose-on-Demand workflow offers a new concept of Personalized Dose Preparation which is currently not possible with a conventional cyclotron.
    کلید واژگان
    Cyclotron
    Positron Emission Tomography Computed Tomography
    Fluorodeoxyglucose 18F

    شماره نشریه
    4
    تاریخ نشر
    2019-08-01
    1398-05-10
    ناشر
    Shiraz University of Medical Sciences
    سازمان پدید آورنده
    PET/CT Imaging Services, Department of Radiology, Aga Khan University Hospital, Karachi, Pakistan
    PET/CT Imaging Services, Department of Radiology, Aga Khan University Hospital, Karachi, Pakistan
    PET/CT Imaging Services, Department of Radiology, Aga Khan University Hospital, Karachi, Pakistan
    Allmed Solution, Karachi, Pakistan
    PET/CT Imaging Services, Department of Radiology, Aga Khan University Hospital, Karachi, Pakistan
    PET/CT Imaging Services, Department of Radiology, Aga Khan University Hospital, Karachi, Pakistan

    شاپا
    2251-7200
    URI
    https://dx.doi.org/10.31661/jbpe.v0i0.792
    https://jbpe.sums.ac.ir/article_46586.html
    https://iranjournals.nlai.ir/handle/123456789/26517

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب