    Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol

    (ندگان)پدیدآور
    Nainggolan, APawiro, S A
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    1.785 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Research
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to create local treatment plans of VMAT and IMRT. The local planning results were compared with the results of TG 119. Point measurement at high and low dose regions were measured using three ionization chambers with different active volumes (CC01, CC13, FC65G). The composite dose was measured by a 2D detector array and analyzed for the percentage of points passing the gamma criteria of 3 % dose difference (DD) and 3 mm distance-to-agreement (DTA) and 2 % DD and 2 mm DTA. Results: The local treatment plans of VMAT and IMRT capable to meet the dose goals criteria set by TG 119 except for C-shape hard. Three ionization chambers with various active volumes for point measurement showed an increase in the confidence limit (CL), the larger the active volume was found proportional to increase the value of CL. The results obtained from ion chambers CC01 and CC13 could met the dose criteria set by TG 119, but results obtained from ion chamber FC65G fail the criteria. All gamma evaluation results show more than 95% data points pass the criteria of 3% DD and 3 mm DTA and the gamma index CL results fall within the TG 119 criteria, which is below 12.4. Conclusion: TG 119 methodology and recommendations have successfully been used to evaluate commissioning accuracy of VMAT. The CL value of the study could be used as a reference and recommendation to evaluate the accuracy and integrity of treatment planning and treatment delivery systems of VMAT and IMRT.
    کلید واژگان
    Radiotherapy, Intensity-Modulated
    Radiometry
    Phantoms
    Imaging Radiotherapy Planning
    Computer-Assisted

    شماره نشریه
    4
    تاریخ نشر
    2019-08-01
    1398-05-10
    ناشر
    Shiraz University of Medical Sciences
    سازمان پدید آورنده
    Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Depok, 16424
    Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Depok, 16424

    شاپا
    2251-7200
    URI
    https://dx.doi.org/10.31661/jbpe.v0i0.839
    https://jbpe.sums.ac.ir/article_46587.html
    https://iranjournals.nlai.ir/handle/123456789/26516

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب