      MRS Shimming: An Important Point Which Should not be Ignored

      (ندگان)پدیدآور
      Zeinali-Rafsanjani, BFaghihi, RMosleh-Shirazi, M ASaeedi-Moghadam, MLotfi, MJalli, RSina, SMina, L
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      1.320 مگابایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Research
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Introduction: Proton magnetic resonance spectroscopy (MRS) is a well-known device for analyzing the biological fluids metabolically. Obtaining accurate and reliable information via MRS needs a homogeneous magnetic field in order to provide well-defined peaks and uniform water suppression. There are lots of reasons which can disturb the magnetic field homogeneity which can be corrected by a process known as shimming. This study is intended to recall the importance of shimming and also the significant role of quality control (QC) in achieving an accurate quantification.Material and Method: An acrylic cylindrical quality control phantom was designed as an analog of brain MRS test phantoms in order to control the accuracy of the obtained signal of a 1.5 T Siemens MRI system which belonged to one of Shiraz hospitals. The signal of NAA, Cho, Cr, the combination of these metabolites and also the distilled water, which was used in this study, was evaluated using separate phantoms. A QC test was performed using Siemens QC phantom and a standard test phantom.Results: The spectrum of our home- made phantom had a significant difference with the expected spectrum. The results of checking the spectrum of metabolites separately also confirmed that there was a systemic problem that affects all the signals originated from all metabolites and even the pure distilled water. The MRS system could not pass QC tests, and peak broadening was common in all spectra. The complex spectrum of standard test phantom was not produced successfully by the MRS system.Discussion: By a simple check of the water peak characteristics, lots of information can be obtained, one of which is the status of shimming that has a considerable effect on the accuracy of the spectrum. Thus, performing an automatic or manual shimming is not a criterion of the spectrum accuracy, and performing a periodic quality control using a test phantom by a specialist  is necessary.Conclusion: Briefly, the quality control of MRS and all the other clinical device must be taken seriously. Sometimes QC can be the boundary of a right or a wrong decision for the patient.
      کلید واژگان
      Magnetic resonance spectroscopy
      Shimming
      quality control
      Quality assurance

      شماره نشریه
      3
      تاریخ نشر
      2018-09-01
      1397-06-10
      ناشر
      Shiraz University of Medical Sciences
      سازمان پدید آورنده
      Department of Nuclear Engineering, School of Mechanical Engineering, Shiraz University, Shiraz, Iran
      Department of Nuclear Engineering, School of Mechanical Engineering, Shiraz University, Shiraz, Iran
      Medical imaging research center, Shiraz University of medical sciences, Shiraz, Iran
      Medical imaging research center, Shiraz University of medical sciences, Shiraz, Iran
      Medical imaging research center, Shiraz University of medical sciences, Shiraz, Iran
      Medical imaging research center, Shiraz University of medical sciences, Shiraz, Iran
      Radiation research center, School of Mechanical Engineering, Shiraz University, Shiraz, Iran
      Radiology department, Faghihi hospital, Shiraz University of medical sciences, Shiraz, Iran

      شاپا
      2251-7200
      URI
      https://dx.doi.org/10.31661/jbpe.v0i0.845
      https://jbpe.sums.ac.ir/article_43320.html
      https://iranjournals.nlai.ir/handle/123456789/26435

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب