      Evaluation of Lung Density and Its Dosimetric Impact on Lung Cancer Radiotherapy: A Simulation Study

      (ندگان)پدیدآور
      Raj Verma, TKumar Painuly, NPrasad Mishra, SYoganathan, S ASingh, NBhatt, M L BJamal, N
      Thumbnail
      دریافت مدرک مشاهده
      FullText
      اندازه فایل: 
      805.4کیلوبایت
      نوع فايل (MIME): 
      PDF
      نوع مدرک
      Text
      Original Research
      زبان مدرک
      English
      نمایش کامل رکورد
      چکیده
      Background: The dosimetric parameters required in lung cancer radiation therapy are taken from a homogeneous water phantom; however, during treatment, the expected results are being affected because of its inhomogeneity. Therefore, it becomes necessary to quantify these deviations.Objective: The present study has been undertaken to find out inter- and intra- lung density variations and its dosimetric impact on lung cancer radiotherapy using Monte Carlo code FLUKA and PBC algorithms.Material and Methods: Density of 100 lungs was recorded from their CT images along with age. Then, after PDD calculated by FLUKA MC Code and PBC algorithm for virtual phantom having density 0.2 gm/cm3 and 0.4 gm/cm3 (density range obtained from CT images of 100 lungs) using Co-60 10 x10 cm2 beams were compared.Results: Average left and right lung densities were 0.275±0.387 and 0.270±0.383 respectively. The deviation in PBC calculated PDD were (+)216%, (+91%), (+)45%, (+)26.88%, (+)14%, (-)1%, (+)2%, (-)0.4%, (-)1%, (+)1%, (+)4%, (+)4.5% for 0.4 gm/cm3 and (+)311%, (+)177%, (+)118%, (+)90.95%, (+)72.23%, (+)55.83% ,(+)38.85%, (+)28.80%, (+)21.79%, (+)15.95%, (+)1.67%, (-) 2.13%, (+)1.27%, (+)0.35%, (-)1.79%, (-)2.75% for 0.2 gm/cm3 density mediums at depths of 1mm, 2mm, 3mm, 4mm, 5mm, 6 mm, 7 mm, 8mm, 9mm,10mm, 15mm, 30mm, 40mm, 50mm, 80mm and 100 mm, respectively.Conclusion: Large variations in inter- and intra- lung density were recorded. PBC overestimated the dose at air/lung interface as well as inside lung. The results of Monte Carlo simulation can be used to assess the performance of other treatment planning systems used in lung cancer radiotherapy.
      کلید واژگان
      PBC
      Monte Carlo Code FLUKA
      Variation in Lung Density
      Virtual Phantom
      Computed Tomography

      شماره نشریه
      1
      تاریخ نشر
      2019-02-01
      1397-11-12
      ناشر
      Shiraz University of Medical Sciences
      سازمان پدید آورنده
      King George Medical University, UP, Lucknow, India
      King George Medical University, UP, Lucknow, India
      Dr.Ram Manohar Lohia Institute of Medical Sciences, Lucknow, India
      Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, India
      King George Medical University, UP, Lucknow, India
      King George Medical University, UP, Lucknow, India
      King George Medical University, UP, Lucknow, India

      شاپا
      2251-7200
      URI
      https://dx.doi.org/10.31661/jbpe.v9i1Feb.430
      https://jbpe.sums.ac.ir/article_43375.html
      https://iranjournals.nlai.ir/handle/123456789/26344

      مرور

      همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

      حساب من

      ورود به سامانهثبت نام

      تازه ترین ها

      تازه ترین مدارک
      © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
      تماس با ما | ارسال بازخورد
      قدرت یافته توسطسیناوب