Chromium (III) ion selective electrode based on di(benzylamino)glyoxime
(ندگان)پدیدآور
Kia, M.Aghaie, H.Arvand, M.Zare, K.Aghaie, M.نوع مدرک
TextResearch Paper
زبان مدرک
Englishچکیده
A new poly(vinylchloride) membrane sensor for Cr3+ ions based on di (benzylamino)glyoxime asan ionophore was prepared. The electrode has a linear dynamic range 1×10-6-1×10-1 mol l-1,with aNernstian slope of 20.3 ± 0.5 mV per decade and a detection limit of 2×10-7. It has a fast responsetime of potential. The proposed electrode revealed good sensitivity for Cr(III) over a wide variety of metalions and could be used in a pH range of 2-5.5. The proposed sensor was used for the determinationof Cr3+ ions in real samples such as tea leaves, coffee and cacao powder by direct potentiometry.