Application of a dioxime-PVC electrode to potentiometric studies of Cr(III) ion
(ندگان)پدیدآور
Aghaie, H.Felcri, M. H.Arvand, M.Zare, K.Aghaie, M.نوع مدرک
TextResearch Paper
زبان مدرک
Englishچکیده
A PVC membrane Cr(III) ion selective electrode has been constructed using 2H-1,4-benzothioazine-2,3(4H)dione dioxime (BTD) as membrane carrier. The influence of membrane composition on the electroderesponse was studied. The electrode exhibits a Nernstian response over a Cr(III) concentration range of 1.0 x10 6 to 1.0>of 8.9 x le M. The selectivity coefficient values as determined by separate solution method (SSM) indicateexcellent selectivity for Cr(III) ion over a large number of other ions. Applications of the proposed electrode tothe determination of Cr(III) in real samples and as an indicator for potentiometric titration of Cr(III) ion withEDTA, are reported.
کلید واژگان
Ion selective electrodePVC membrane
Cr(III) determination
a-dioxime
شماره نشریه
3تاریخ نشر
2006-11-011385-08-10
ناشر
Tehran, Islamic Azad University of Iran, Science and Research Branchسازمان پدید آورنده
Not-mentionedNot-mentioned
Not-mentioned
Not-mentioned
Not-mentioned