A Novel Potentiometric Method for the Determination of Atenolol in Pure and Commercial Products
(ندگان)پدیدآور
Katif, HasanHassan, Omar
نوع مدرک
TextOriginal Article
زبان مدرک
Englishچکیده
To ascertain the amount of atenolol present in pharmaceutical and pure samples, two distinct kinds of ion selective electrodes were developed: coated graphite electrodes (CGE) and coated wire electrodes (CWE). At the range 6.2×10-6-1.0×10-2 and 4.5×10-8-1.0×10-2 mol.L-1 with a slope of 52.95, and 56.23 mV/decade with a low limit of detection 1.8×10-6 and 1.3×10-8 mol.L-1, the electrodes adhered to the Nernst equation for CGE and CWE electrodes, respectively. Additionally, the CWE and CGE electrodes had response times of 38 and 26 seconds and life times of 34 and 41 days, respectively. Moreover, the recommended sensors showed long-term stability and functioned after preparation with negligible variations in the potential or slope values. In this study, the ATL drug's high selectivity over a number of inorganic cations was demonstrated by the ATL sensors that were made for the drug formulation samples. The sensors being studied allow for the direct determination of ATL in specific Commercial Products.
کلید واژگان
AtenololISEs
Direct method
Nernstian response
Selectivity
Coated wire and graphite electrode
Potentiometry
شماره نشریه
4تاریخ نشر
2025-04-011404-01-12
ناشر
University of Tehranسازمان پدید آورنده
Department of Chemistry, College of Education for Pure Science, Tikrit University, Tikrit, IraqDepartment of Chemistry, College of Education for Pure Science, Tikrit University, Tikrit, Iraq