    Is a Complete Anatomical Fit of the Tomofix Plate Biomechanically Favorable? A Parametric Study Using the Finite Element Method

    (ندگان)پدیدآور
    Hayatbakhsh, ZahraFarahmand, FarzamKarimpour, Morad
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    852.5کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    RESEARCH PAPER
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Background: The opening wedge high tibial osteotomy (HTO) fixation using the Tomofix system is at the risk of mechanical failure due to unstable fixation, lateral hinge fracture, and hardware breakage. This study aimed to investigate the effect of the level of anatomical fit (LOF) of the plate on the failure mechanisms of fixation.Methods: A finite element model of the HTO with a correction angle of 12 degrees was developed. The LOF of the TomoFix plate was changed parametrically by altering the curvature of the plate in the sagittal plane. The effect of the LOF on the fixation performance was studied in terms of the factor of safety (FOS) against failure mechanisms. The FOSs were found by 1) dividing the actual stiffness of the plate-bone construct by the minimum allowable one for unstable fixation, 2) dividing the compressive strength of the cortical bone by the actual maximum pressure at the lateral hinge for the lateral hinge fracture, and 3) the Soderberg criterion for fatigue fracture of the plate and screws.Results: The increase of the LOF by applying a larger bent to the plate changed the fixation stiffness slightly. However, it reduced the lateral hinge pressure substantially (from 182 MPa to 71 MPa) and increased the maximum equivalent stresses in screws considerably (from 187 MPa to 258 MPa). Based on the FOS-LOF diagram, a gap smaller than 2.3 mm was safe, with the highest biomechanical performance associated with a 0.5 mm gap size.Conclusion: Although a high LOF is necessary for the Tomofix plate fixation to avoid mechanical failure, a gap size of 0.5mm is favored biomechanically over complete anatomical fit. Level of evidence: V
    کلید واژگان
    Failure mechanisms
    Fixation stiffness
    Hardware breakage
    Lateral hinge pressure
    Plate contouring

    شماره نشریه
    8
    تاریخ نشر
    2022-08-01
    1401-05-10
    ناشر
    Mashhad University of Medical Sciences, Iranian Society of Knee Surgery, Arthroscopy and Sports Tramatology,Iranian Orthopaedic Association
    سازمان پدید آورنده
    Department of Biomedical Engineering, Science and Research branch, Islamic Azad University, Tehran, Iran
    Mechanical Engineering Department, Sharif University of Technology, Tehran, Iran
    School of Mechanical Engineering, University of Tehran, Tehran, Iran

    شاپا
    2345-4644
    2345-461X
    URI
    https://dx.doi.org/10.22038/abjs.2022.60928.3003
    https://abjs.mums.ac.ir/article_20587.html
    https://iranjournals.nlai.ir/handle/123456789/927600

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب