    A Comparison of Patients Absorption Doses with Bone Deformity Due to the EOS Imaging and Digital Radiology

    (ندگان)پدیدآور
    Abrisham, Seyed Mohammad J.Bouzarjomehri, FathollahNafisi-Moghadam, RezaSobhan, Mohammad R.Gadimi, MahdieOmidvar, Fereshte
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    346.1کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    RESEARCH PAPER
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
      Background: This study has aimed to measure the patient dose in entire spine radiography by EOS system in comparison with the digital radiography. Methods: EOS stereo-radiography was used for frontal and lateral view spine imaging in 41 patients in a prospective analytical study. A calibrated dose area product (DAP) meter was used for calibration of the DAP in EOS system. The accuracy and precision of the system was confirmed according to the acceptance testing. The same procedure was used for 18 patients referred for lumbar spine digital radiology (overall 36 images). Results: Although radiation fields in the EOS were almost twice of that in digital radiology, and the average peak tube voltage (kVp), current supply to the tube (mA), and the average size and age of the patients referred for EOS imaging were greater than digital radiology, however, the average DAP in EOS was 1/5 of that in digital radiology system. Also, the average dose in the EOS was about 1/20 of that in digital radiology. Conclusion: The patient dose in EOS imaging system was lower in comparison with digital radiology (1/20).
    کلید واژگان
    Digital Radiography
    Dose Area Product
    Dosimetry
    EOS
    Stereo Radiography
    General Orthopedics

    شماره نشریه
    3
    تاریخ نشر
    2017-05-01
    1396-02-11
    ناشر
    Mashhad University of Medical Sciences, Iranian Society of Knee Surgery, Arthroscopy and Sports Tramatology,Iranian Orthopaedic Association
    سازمان پدید آورنده
    Department of Orthopedics, Shahid Sadoughi University of Medical Sciences, Yazd, Iran
    Medical Radiation group, Islamic Azad University, Central Tehran branch, Tehran, Iran
    Department of Radiology, Shahid Sadoughi University of Medical Sciences, Yazd, Iran
    Department of Orthopedics, Shahid Sadoughi University of Medical Sciences, Yazd, Iran
    Department of Medical Physics, Shahid Sadoughi University of Medical Sciences, Yazd, Iran
    Department of Medical Physics, Shahid Sadoughi University of Medical Sciences, Yazd, Iran

    شاپا
    2345-4644
    2345-461X
    URI
    https://dx.doi.org/10.22038/abjs.2017.21918.1568
    http://abjs.mums.ac.ir/article_8599.html
    https://iranjournals.nlai.ir/handle/123456789/431709

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب