    A comparison of 2D and 3D kidney absorbed dose measures in patients receiving 177Lu-DOTATATE

    (ندگان)پدیدآور
    Willowson, KathyRyu, HyunJuJackson, PriceEslick, EnidSingh, AnitaBailey, Dale
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    640.1کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Article
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Objective(s): To investigate and compare quantitative accuracy of kidney absorbed dose measures made from both 2D and 3D imaging in patients receiving 177LuDOTATATE (Lutate) for treatment of neuroendocrine tumours (NETs). Methods: Patients receiving Lutate therapy underwent both whole body planar imaging and SPECT/CT imaging over the kidneys at time points 0.5, 4, 24, and 96-120 hours after injection. Planar data were corrected for attenuation using transmission data, and were converted to units of absolute activity via two methods, using either a calibration standard in the field of view or relative to pre-voiding image total counts. Hand drawn regions of interest were used to generate time activity curves and kidney absorbed dose estimates in OLINDA-EXM. Fully quantitative SPECT data were generated using CT-derived corrections for both scatter and attenuation, before correction for dead time and application of a camera specific sensitivity factor to convert data to units of absolute activity. Volumes of interest were defined for kidney using the co-registered x-ray CT, before time activity curves and absorbed dose measures were generated in OLINDA-EXM, both with and without corrections made to the model for patient specific kidney volumes. Quantitative SPECT data were also used to derive dose maps through dose kernel convolution (DKC), which was treated as the gold standard. Results: A total of 50 studies were analysed, corresponding to various cycles of treatment from 21 patients. Planar absorbed dose estimates were consistently higher than SPECT derived estimates by, on average, a factor of 3. Conclusion: Quantitative SPECT is considered the gold standard approach for organ specific dosimetry however often relies on in house software. As such planar methods for estimating absorbed dose are much more widely available, and in particular, are often the only source of reference in previously published data. For the case of Lutate dosimetry, planar measures may lead to a three-fold increase in measures of kidney absorbed dose.
    کلید واژگان
    Lutate
    Dosimetry
    Kidney
    NET
    Image Quantification

    شماره نشریه
    2
    تاریخ نشر
    2018-07-01
    1397-04-10
    ناشر
    Asia Oceania Federation of Nuclear Medicine & Biology
    سازمان پدید آورنده
    Institute of Medical Physics, The University of Sydney
    Faculty of Health Sciences, University of Sydney, Lidcombe 2141, Australia
    Department of Cancer Imaging, Peter MacCallum Cancer Centre, Melbourne 3000, VIC, Australia
    Institute of Medical Physics, School of Physics, University of Sydney, 2006 NSW, Australia
    Department of Nuclear Medicine, Royal North Shore Hospital, St Leonards 2065, NSW, Australia
    Department of Nuclear Medicine, Royal North Shore Hospital, St Leonards 2065, NSW, Australia Faculty of Health Sciences, University of Sydney, Lidcombe 2141, Australia

    شاپا
    2322-5718
    2322-5726
    URI
    https://dx.doi.org/10.22038/aojnmb.2018.26105.1182
    http://aojnmb.mums.ac.ir/article_10329.html
    https://iranjournals.nlai.ir/handle/123456789/420978

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب