Effect of Recycling and Autoclave Sterilization on the Unloading Forces of NiTi Closed-Coil Springs: An In Vitro Study
(ندگان)پدیدآور
Momeni Danaei, Sh.Oshagh, M.Khozaei, A.
نوع مدرک
TextOriginal Article
زبان مدرک
Englishچکیده
Statement of Problem: Clinicians use the NiTi coil springs frequently for its appropriate mechanical properties.Purpose: The aim of this study was to determine the effect of recycling and autoclave sterilization on the unloading forces of NiTi closed coil springs.Materials and Method: Fourteen NiTi closed coil spring with the length of 9mm were selected. Each coil was stretched to a peak extension of 12 mm. A universal testing machine was used to acquire load/deflection curve of the coil springs at 25±2°C. The influence of thermocycling (1000 cycles,5-55°C), autoclaving (134°C, 32PSI, 3min) and mechanical strain (9mm extension) which simulated the oral condition, were considered. Data were statistically analyzed by adopting Repeated Measures MANOVA Paired t-Test.Results: Autoclaving in the 1, 4, 6 steps increased the force levels of coil springs about 2-5gf (p