    Impact of Silicone gel breast prosthesis on photon beam therapy of recurrence carcinoma post mastectomy

    (ندگان)پدیدآور
    Letsoalo, RamaeselaSithole, MphoNyathi, Mpumelelo
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    954.2کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    AbstractIntroduction Silicone gel breast implants are used for breast reconstruction post mastectomy. In the event of cancer recurrence, the radiation oncologists are forced to irradiate through the prosthesis device. Due to prosthesis higher atomic number dose perturbations occur during treatment. This study determined the influence of silicone gel thickness on the photon beam distribution. Materials and MethodsA Varian linear accelerator, water phantom (dimensions 30 × 30 × 30 cm3), silicone gel breast prosthesis and Omni-Pro Accept software were used in the study. With the gantry positioned at at a source-to-surface distance (SSD) of 100 cm, the collimators were adjusted to a field size of 10 × 10 cm2 and the 6 MeV photon beam was used. Omni-Pro Accept software was used to plot the percentage depth dose curves and the beam profiles. The results obtained with a Monte Carlo Nuclear Particle (MCNP) Code were validated with the measurement data. The beam profile and percentage depth dose curves were also measured for silicone gel thicknesses (4, 6, 8, 10, 12, 14 and 16 cm) and width 16.5 cm aligned 1 cm below the surface of the water.ResultsThe measured and calculated percentage depth dose (PDD) ratio was 0.03. The measured and calculated beam profiles were 0.5% and 0.9% respectively. For the silicone gel prosthesis, the depth dose values at 0.5 cm below the prosthesis were 2.8%.ConclusionDose perturbations below the breast prosthesis are insignificant, breast prosthesis are safe in the event of carcinoma recurrences.
    کلید واژگان
    Keywords: Monte Carlo method
    Silicone gel breast prosthesis
    Photon therapy
    Photon dose distribution, MCNP code, Percentage depth dose
    Medical Application of Computer Simulation
    Medical Physics
    Physics of Radiotherapy
    Radiation Detection & Measurement

    شماره نشریه
    4
    تاریخ نشر
    2018-10-01
    1397-07-09
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Department of Medical Physics, Faculty of Medicine, Sefako Makgatho Health Sciences University, Pretoria, South Africa
    Department of Physics, Faculty of , Sefako Makgatho Health Sciences University, Pretoria, South Africa
    Department of Medical Physics Sefako Makgatho Health Sciences University SouthAfrica

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2018.27148.1281
    http://ijmp.mums.ac.ir/article_10514.html
    https://iranjournals.nlai.ir/handle/123456789/324916

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب