Assessment of Patient Radiation Dose in Interventional Procedures at Shahid Madani Heart Center in Khorramabad, Iran
(ندگان)پدیدآور
Gholami, Mehrdadgharloghi, SomayehZare, SoodabehSaki, ArezooPiri, ZahraMousavi, Majidنوع مدرک
TextOriginal Paper
زبان مدرک
Englishچکیده
Introduction: Coronary angiography is the most common angiographic procedure for diagnosis and treatment of the heart diseases. Herein, we aimed to evaluate the entrance surface dose (ESD), dose area product (DAP), as well as cancer risk in interventional cardiology procedures. Materials and Methods: This study was conducted during July-December 2015 at Shahid Madani Heart Center in Khorramabad, Iran. A total of 225 adult patients including 122 females and 103 males regardless of the risk factors for coronary diseases were participated. Of them, 199 and 26 patients underwent diagnostic coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA), respectively. Each patient underwent CA or PTCA separately. All the procedures were carried out using Siemens angiography system with the pulsed fluoroscopy of 10-30 pulses/s and cine frame rate of 15 frames/s. DAP, ESD, fluoroscopy time (FT), as well as the number of sequences and frames per sequence were collected for each 199 CA and 26 PTCA procedures. Results: The median values of DAP were 19.77±14.88 and 57.11±33.36 Gy.cm2 in CA and PTCA, respectively. In addition, the median values of ESD were 323.12±245.39 and 1145.22±594.42 mGy in CA and PTCA, respectively. FTs were 114.59±74.33 s in CA and 424.15±292.93 s in PTCA. Conclusion: The average patient dose and cancer risk estimates in both CA and PTCA were consistent with the reference levels. However, in agreement with other interventional procedures, dose levels in the interventional cardiology are influenced by staff and clinical protocols, as well as the type of equipment.
کلید واژگان
Coronary Angiography Entrance Surface DoseDose Area Product
Effective Dose
Radiation risk
Medical Physics
Physics of Diagnostics Radiology
Radiation Detection & Measurement
شماره نشریه
3تاریخ نشر
2017-09-011396-06-10
ناشر
Mashhad University of Medical Sciencesسازمان پدید آورنده
Department of Medical Physics & Radiation Sciences, Lorestan University of Medical Sciences, Khorram Abad, IranDepartment of Medical physics and Radiation Sciences, School of Para Medicine, Lorestan University of Medical Sciences, Khorramabad, Iran
Department of Public Health, Lorestan University of Medical Sciences, Khorramabad, Iran
Student Research Committee, Lorestan University of Medical Sciences, Khorramabad, Iran
Student Research Committee, Lorestan University of Medical Sciences, Khorramabad, Iran
Angiography and Angioplasty section, Shahid Madani Heart Center, Lorestan University of Medical Sciences, Khorramabad, Iran