    Dosimetric Effects of Rotational Setup Error in Volumetric Modulated Arc Radiotherapy on Brain Tumor Patients

    (ندگان)پدیدآور
    Park, Hyo-KukCho, Jae-HwanKim, sungchul
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    1.165 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: This study examined the dosimetric effects based on the rotational setup error to correct patient setup errors occur during volumetric modulated arc radiotherapy (VMAT) for brain tumor patients. Material and Methods: This study included 1129 cases of cone beam computed tomography (CBCT) images obtained from 46 brain tumor patients, who experienced VMAT and used the 6DoF (degree of freedom) treatment couch. The dosimetric effects regarding the application of the rotational setup error were examined by comparing the treatment plans. Results: The mean patient setup errors at the lateral (X-axis), longitudinal (Y-axis), and vertical (Z-axis) directions were 0.1±1.4, 0.0±1.1, and -0.4±1.2 mm, respectively. The pitch, roll, and yaw were -0.29±0.61°,-0.42±0.98°, and -0.53±0.69°, respectively. When an absolute value was taken for the setup error, the mean error was 1.06±0.14 mm at the three translation directions, and the error of rotation was 0.82±0.14°, showing a larger error than that of translation. In terms of the mean dose difference by each region of interest (ROI) before and after correcting for the rotational setup error on the treatment plan, Brain_max was 2.17 Gy, and Brain_mean was 0.28 Gy, whereas the maximum and mean of Brain_stem were -3.58 and -4.43 Gy, respectively. These findings suggested a dose difference according to the correction of the rotational setup error. Conclusion: This study indicated that the dose effect is influenced by the rotational setup error in VMAT of brain tumor patients. Moreover, the 6DoF positional correction could reduce the positional uncertainties and deliver a more accurate dose.
    کلید واژگان
    Volumetric modulated arc radiotherapy
    Cone beam computed tomography
    Setup error
    Dose
    Medical Physics
    Physics of Radiotherapy

    شماره نشریه
    2
    تاریخ نشر
    2019-03-01
    1397-12-10
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Department of Radiation Oncology, Yonsei Cancer Center, Seoul
    Department of Radiological Technology, Ansan University, Ansan
    Radiological Science, Gachon University Medical Campus, Incheon, Korea

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2018.32544.1392
    http://ijmp.mums.ac.ir/article_11425.html
    https://iranjournals.nlai.ir/handle/123456789/324825

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب