• ثبت نام
    • ورود به سامانه
    مشاهده مورد 
    •   صفحهٔ اصلی
    • نشریات انگلیسی
    • Iranian Journal of Medical Physics
    • Volume 10, Issue 2
    • مشاهده مورد
    •   صفحهٔ اصلی
    • نشریات انگلیسی
    • Iranian Journal of Medical Physics
    • Volume 10, Issue 2
    • مشاهده مورد
    JavaScript is disabled for your browser. Some features of this site may not work without it.

    Calculations of Linac Photon Dose Distributions in Homogeneous Phantom Using Spline

    (ندگان)پدیدآور
    Sadoughi, Hamid-RezaNasseri, ShahrokhMomennezhad, MahdiSadoghi-Yazdi, HadiZare, Mohammad-HosseinBahreyni-Toosi, Mohammad-Hossein
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    499.8کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction Relative dose computation is a necessary step in radiation treatment planning. Therefore, finding an approach that is both fast and accurate seems to be necessary. The purpose of this work was to investigate the feasibility of natural cubic spline to reconstruct dose maps for linear accelerator radiation treatment fields in comparison with those of the simulation. Materials and Methods A natural cubic spline algorithm was used to reproduce dose calculations of linac radiation treatment fields resulting from GEANT4 application for tomographic emission (GATE) simulation. The spline algorithm was used to compute percent depth dose of radiation therapy fields for 6 MV X-rays, which were calculated by simulation of Elekta Compact Linac. It reconstructed 2-dimensional dose maps and created isodose distributions. This dose maps were evaluated and compared with the simulation, where the γ -index was used. Results A good agreement was found between the doses calculated from the simulation and the spline. In particular, an average γ-index passing rate of 0.24 was obtained for sample percent depth dose distributions, and an average γ -index passing rate of 0.20 was observed for sample dose profiles. Conclusion Natural cubic spline has been established to calculate dose maps from field characteristics. The feasibility and possibility of natural cubic spline to calculate dose maps for linac radiation therapy fields in a homogeneous phantom has been demonstrated.
    کلید واژگان
    γ –index
    Cubic Spline
    Dose Calculation
    Medical Physics
    Physics of Radiotherapy
    Radiation Detection & Measurement

    شماره نشریه
    2
    تاریخ نشر
    2013-09-01
    1392-06-10
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Medical Physics Research Center, Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran
    Medical Physics Research Center, Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran
    Medical Physics Research Center, Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran Imam Reza Hospital, Mashhad University of Medical Sciences, Mashhad, Iran
    Department of Computer Engineering, Ferdowsi University of Mashhad, Mashhad,
    Medical Physics Research Center, Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran
    Medical Physics Research Center, Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2013.2181
    http://ijmp.mums.ac.ir/article_2181.html
    https://iranjournals.nlai.ir/handle/123456789/324814