    Establishment of Diagnostic Reference Levels and Estimation of Effective Dose from Computed Tomography Head Scans at a Tertiary Hospital in South Africa

    (ندگان)پدیدآور
    Nyathi, Mpumelelo
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    1.150 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: Head scans are the most frequently performed computed tomography (CT) examinations worldwide. However, there is growing concern over the probability of increased cancer risks among the exposed populations. Diagnostic reference levels (DRLs) identify radiation dose that is not commensurate with clinical objectives. The aim of this study was to establish DRLs for CT head procedures and estimate effective dose (ED). Material and Methods: The dose absorbed by the head slice of a Rando Alderson phantom was measured using calibrated lithium fluoride thermoluminescent dosimeters (TLDs) exposed to a CT scanner operated on clinical parameters. The measurements were done at the periphery and center of the slice, and repeated twice with a new set of TLDs. The radiation dose absorbed by the TLDs was read using a Harshaw TLD reader, Model 5500. The measured doses were used to calculate the weighted CT dose index (CTDIw), CT dose index volume (CTDIv), and dose length product (DLP). Finally, the ED was calculated using the formula; ED = k × DLP, where k was considered as 0.0021. Results: The mean absorbed dose was 30.9 mGy, while the established CTDIv and DLP values for the head protocol were 40 mGy and 990 mGy.cm, respectively. Additionally, the ED was calculated as 2.1 mSv. These values compared well with some international values. Conclusion: According to the results of the present study, the established CTDIv, DLP, and ED for head scan were well-compared with some international values, except in the cases using different scan lengths and scanner algorithms.
    کلید واژگان
    Effective Dose
    Computed Tomography Dose
    Absorbed Dose
    Computed Tomography (CT)
    Medical Physics
    Radiation Detection & Measurement

    شماره نشریه
    2
    تاریخ نشر
    2020-03-01
    1398-12-11
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Department of Medical Physics Sefako Makgatho Health Sciences University SouthAfrica

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2019.39685.1531
    http://ijmp.mums.ac.ir/article_13341.html
    https://iranjournals.nlai.ir/handle/123456789/324788

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب