    A Validation Study on Radiation properties of a Novel LiF: Mg, Ti Known as IAP-100

    (ندگان)پدیدآور
    Mohammadi, KheirollahGholamhosseinian, HamidBaziar, OmidMohammadi, AghilBayani Roodi, Shahram
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    489.3کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: LiF dosimeter has the most application in medicine. This study aimed to evaluate some dosimetric properties of a novel LiF: Mg, Ti. Materials and Methods: An ELEKTA Precise linear accelerator was used to calibrate dosimeters at 6 MV. In this survey, responses of dosimeters were evaluated up to 1000 cGy. Background effect was investigated in two different dosimeter states including irradiated and unirradiated.Thermoluminescence response dependence to dose rate was investigated, as well. Energy dependence was evaluated in diagnostic and therapeutic ranges. Furthermore, fading effect was evaluated by reading the dosimeters every 2 h up to 12 h post-irradiation.  Results: The dosimeters had linear response up to 250 cGy. Readout values of dosimeters receiving 120 cGy at three dose rates of 21, 212, 425 cGy.min-1 were calculated equal to 125, 123, 121 cGy, respectively. The measured values of delivering 80, 120, and 150 cGy prescribed doses at 6 MV, 10 MV, and 15 MV were accurate at 6 MV and about 1.5 times higher than the prescribed doses at 10 and 15 MV. Thermoluminescence response in diagnostic energy range showed an uprising trend with increasing energy. Conclusion: The raising thermoluminescence response with increasing energy contradicts with the findings of Nunn. Due to the reproducibility and linear response of dosimeters in an acceptable dose range, they could be used in diagnostic and therapeutic fields. Effects of absorbed doses from background in low-dose studies, mainly in diagnostic radiology range, could be evaluated in more detail in future surveys.
    کلید واژگان
    Radiation Dosimeters
    Reproducibility of Results
    Reading
    Calibration
    Medical Physics
    Radiation Detection & Measurement

    شماره نشریه
    2
    تاریخ نشر
    2018-04-01
    1397-01-12
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Malekashtar University,Tehran, Iran
    Medical Physics Dept, Faculty of Medicine, Mashhad University of Medical Sciences
    Medical Physics, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran
    Energy Engeineering and physics faculty, Amirkabir University of Technology, Tehran, Iran
    Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2017.24584.1248
    http://ijmp.mums.ac.ir/article_9400.html
    https://iranjournals.nlai.ir/handle/123456789/324782

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب